Suppr超能文献

光子计数 CT:冠状动脉支架的高分辨率成像。

Photon-Counting CT: High-Resolution Imaging of Coronary Stents.

出版信息

Invest Radiol. 2018 Mar;53(3):143-149. doi: 10.1097/RLI.0000000000000420.

Abstract

PURPOSE

The aim of this study was to investigate computed tomography (CT) imaging characteristics of coronary stents using a novel photon-counting detector (PCD) in comparison with a conventional energy-integrating detector (EID).

MATERIALS AND METHODS

In this in vitro study, 18 different coronary stents were expanded in plastic tubes of 3 mm diameter, were filled with contrast agent (diluted to an attenuation of 250 Hounsfield units [HU] at 120 kVp), and were sealed. Stents were placed in an oil-filled custom phantom calibrated to an attenuation of -100 HU at 120 kVp for resembling pericardial fat. The phantom was positioned in the gantry at 2 different angles at 0 degree and 90 degrees relative to the z axis, and was imaged in a research dual-source PCD-CT scanner. Detector subsystem "A" used a standard 64-row EID, while detector subsystem "B" used a PCD, allowing high-resolution scanning (detector pixel-size 0.250 × 0.250 mm in the isocenter). Images were obtained from both detector systems at identical tube voltage (100 kVp) and tube current-time product (100 mA), and were both reconstructed using a typical convolution kernel for stent imaging (B46f) and using the same reconstruction parameters. Two independent, blinded readers evaluated in-stent visibility and measured noise, intraluminal stent diameter, and in-stent attenuation for each detector subsystem. Differences in noise, intraluminal stent diameter, and in-stent attenuation where tested using a paired t test; differences in subjective in-stent visibility were evaluated using a Wilcoxon signed-rank test.

RESULTS

Best results for in-stent visibility, noise, intraluminal stent diameter, and in-stent attenuation in EID and PCD were observed at 0-degree phantom position along the z axis, suggesting higher in-plane compared with through-plane resolution. Subjective in-stent visibility was superior in coronary stent images obtained from PCD compared with EID (P < 0.001). Mean in-stent diameter was 28.8% and 8.4% greater in PCD (0.85 ± 0.24 mm; 0.83 ± 0.14 mm) as compared with EID acquisitions (0.66 ± 0.21 mm; 0.76 ± 0.13 mm) for both 0-degree and 90-degree phantom positions, respectively. Average noise was significantly lower (P < 0.001) for PCD (5 ± 0.2 HU) compared with EID (8.3 ± 0.2 HU). The increase in in-stent attenuation (0 degree: Δ 245 ± 163 HU vs Δ 156.5 ± 126 HU; P = 0.006; 90 degrees: Δ 194 ± 141 HU vs Δ 126 ± 78 HU; P = 0.001) was significantly lower for PCD compared with EID acquisitions.

CONCLUSIONS

At matched CT scan protocol settings and identical image reconstruction parameters, the PCD yields superior in-stent lumen delineation of coronary artery stents as compared with conventional EID arrays.

摘要

目的

本研究旨在使用新型光子计数探测器(PCD)与传统能量积分探测器(EID)比较,研究冠状动脉支架的计算机断层扫描(CT)成像特征。

材料和方法

在这项体外研究中,将 18 种不同的冠状动脉支架扩张在直径为 3 毫米的塑料管中,并用对比剂填充(在 120 kVp 时稀释至 250 亨氏单位[HU]的衰减),并密封。将支架放置在一个充满油的定制体模中,在 120 kVp 时校准为-100 HU 的衰减,以模拟心包脂肪。体模以 0 度和 90 度相对于 z 轴的 2 个不同角度放置在机架中,并在研究用双源 PCD-CT 扫描仪中进行成像。探测器子系统“ A”使用标准的 64 排 EID,而探测器子系统“ B”使用 PCD,允许高分辨率扫描(在等中心处探测器像素尺寸为 0.250×0.250 毫米)。从两个探测器系统在相同的管电压(100 kVp)和管电流-时间乘积(100 mA)下获得图像,并均使用典型的支架成像卷积核(B46f)和相同的重建参数进行重建。两名独立的、盲目的读者评估了每个探测器子系统的支架内可视性和测量噪声、管腔内支架直径和支架内衰减。使用配对 t 检验测试噪声、管腔内支架直径和支架内衰减的差异;使用 Wilcoxon 符号秩检验评估主观支架内可视性的差异。

结果

在 z 轴上沿 0 度体模位置观察到 EID 和 PCD 的支架内可视性、噪声、管腔内支架直径和支架内衰减的最佳结果,表明与平面内相比具有更高的平面分辨率。与 EID 相比,PCD 获得的冠状动脉支架图像的主观支架内可视性更优(P<0.001)。0 度和 90 度体模位置的 PCD 采集的支架直径分别比 EID 采集的支架直径大 28.8%和 8.4%(分别为 0.85±0.24 毫米和 0.83±0.14 毫米)。平均噪声显著低于 PCD(5±0.2 HU)(P<0.001)比 EID(8.3±0.2 HU)。与 EID 采集相比,支架内衰减的增加(0 度:Δ245±163 HU 与 Δ156.5±126 HU;P=0.006;90 度:Δ194±141 HU 与 Δ126±78 HU;P=0.001)在 PCD 中明显较低。

结论

在匹配的 CT 扫描方案设置和相同的图像重建参数下,PCD 与传统的 EID 阵列相比,可提供更好的冠状动脉支架管腔描绘。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验