Paradkar Neeraj, Chowdhury Shubhajit Roy
Annu Int Conf IEEE Eng Med Biol Soc. 2017 Jul;2017:113-116. doi: 10.1109/EMBC.2017.8036775.
Cardiovascular Diseases (CVDs) cause a very large number of casualties around the world every year and cardiac arrhythmias contribute to significant proportion of CVD related deaths. Bedside cardiac activity monitors in hospitals are based on electrocardiogram (ECG) processing and are known to produce too many false alarms. Moving beyond bedside care, ECG is not very suitable for use in wearable devices. Photoplethysmography (PPG) on the other hand provides an inexpensive and more wearable device-friendly alternative. This work presents a technique to detect life threatening arrhythmias using only PPG waveforms. PhysioNet Challenge 2015 data is used to detect five types of arrhythmias namely, tachycardia, bradycardia, asystole, ventricular tachycardia and ventricular fibrillation. A novel technique is employed to assign pulse quality index to every PPG pulse and highest quality portion of the signal is used for detection. Results indicate that PPG provides a viable alternative for conventional ECG based detection. An overall true positive rate (TPR) of 93% was achieved with true negative rate (TNR) of 53.78% suggesting that PPG is a viable option for arrhythmia detection.
心血管疾病(CVDs)每年在全球造成大量人员伤亡,而心律失常在与CVD相关的死亡中占很大比例。医院中的床边心脏活动监测器基于心电图(ECG)处理,并且已知会产生过多的误报。除了床边护理之外,ECG不太适合用于可穿戴设备。另一方面,光电容积脉搏波描记法(PPG)提供了一种价格低廉且对可穿戴设备更友好的替代方案。这项工作提出了一种仅使用PPG波形来检测危及生命的心律失常的技术。使用PhysioNet Challenge 2015数据来检测五种类型的心律失常,即心动过速、心动过缓、心搏停止、室性心动过速和心室颤动。采用了一种新技术为每个PPG脉搏分配脉搏质量指数,并使用信号的最高质量部分进行检测。结果表明,PPG为基于传统ECG的检测提供了一种可行的替代方案。实现了93%的总体真阳性率(TPR)和53.78%的真阴性率(TNR),这表明PPG是心律失常检测的一个可行选择。