Boman Eeva, Paterson Dean, Huang Junou, Pearson Shelley, Johnson Carol
Radiation Treatment Department, Blood and Cancer Centre, Wellington Hospital, Wellington, New Zealand.
J Contemp Brachytherapy. 2017 Oct;9(5):418-423. doi: 10.5114/jcb.2017.70987. Epub 2017 Oct 20.
Nylon 6/6 interstitial brachytherapy catheters may stretch when exposed to moisture, mechanical tension, and body temperature. The purpose of this study is to evaluate the magnitude of catheter stretch during a course of multi-catheter interstitial breast brachytherapy for nylon 6/6 treatment catheters, and to assess the impact this has on treatment plan dosimetry.
Catheters were exposed to water at 37°C for six days and the internal catheter length change () was measured daily. Additionally, the measured internal catheter length change (), performed prior to each treatment fraction, for seven interstitial breast brachytherapy patients (total of 76 catheters) were retrospectively analyzed. The maximum length change seen in and were introduced as a source positional error to retrospective treatment plans, and treatment plan dosimetry was analyzed.
and were on average +4.0% and +1.6% of the catheter lengths in water or tissue after 48 hours. Weak correlation was seen between the average per catheter and both the catheter length within tissue (ρ = 0.36, = 0.0007), and the mid-catheter depth in tissue (ρ = 0.42, < 0.0001). The D decreased 1.5% ( < 0.05) and 8.2% ( < 0.05) when the and were introduced to the initial plans.
Nylon 6/6 catheters stretch during a course of multi-catheter interstitial breast brachytherapy treatment. The observed stretch may affect treatment plan dosimetry, if the catheter internal length is only measured immediately after the insertion. Additional catheter length checks are recommended to verify the actual catheter internal length during the treatment.
尼龙6/6组织间近距离治疗导管在接触水分、机械张力和体温时可能会伸展。本研究的目的是评估在多导管组织间乳腺近距离治疗过程中,尼龙6/6治疗导管的伸展程度,并评估其对治疗计划剂量学的影响。
将导管在37°C的水中暴露6天,每天测量导管内部长度变化()。此外,对7例组织间乳腺近距离治疗患者(共76根导管)在每次治疗分次前测量的导管内部长度变化()进行回顾性分析。将在和中观察到的最大长度变化作为源位置误差引入回顾性治疗计划中,并分析治疗计划剂量学。
48小时后,和在水中或组织中的长度平均分别增加4.0%和1.6%。每根导管的平均与组织内导管长度(ρ = 0.36, = 0.0007)以及组织中导管中点深度(ρ = 0.42, < 0.0001)之间均呈弱相关。当将和引入初始计划时,D分别降低了1.5%( < 0.05)和8.2%( < 0.05)。
在多导管组织间乳腺近距离治疗过程中,尼龙6/6导管会伸展。如果仅在插入后立即测量导管内部长度,观察到的伸展可能会影响治疗计划剂量学。建议在治疗期间进行额外的导管长度检查,以验证实际的导管内部长度。