Suppr超能文献

三焦点人工晶状体植入术中急性混浊:一例报告

Acute clouding of trifocal lens during implantation: a case report.

作者信息

Liu Qian, Zhang Suhua, Wang Xiaogang, Cao Weifang, Jia Yading

机构信息

Shanxi Eye Hospital, No. 100 Fudong Street, Taiyuan, Shanxi, 030002, People's Republic of China.

出版信息

BMC Ophthalmol. 2017 Dec 8;17(1):242. doi: 10.1186/s12886-017-0636-7.

Abstract

BACKGROUND

Intraoperative IOLs clouding of several kinds of hydrophilic acrylic intraocular lenses (IOLs) have been reported due to temperature changes. This phenomenon reported previously occurred in cold countries and during the winter months. However, no clinical case was reported about trifocal IOL opacification during operation. We report a case in which acute opacification of the optical region occurred simultaneously when AT LISA tri 839mp(Carl Zeiss) was implanted into the eye.

CASE PRESENTATION

A 79-year-old woman with a cortex and nucleus cataract was scheduled to undergo right eye phacoemulsification assisted by femtosecond technique. The trifocal lens (AT LISA tri 839mp), which is made of hydrophilic acrylic (25%) with hydrophobic surface properties, was chosen for implantation. As the IOL was implanted into the eye, it became cloudy immediately. Then it was replaced by another AT LISA tri 839mp, which was transferred from lens company outside, the same phenomenon was observed. These two lenses underwent the same temperature fluctuation from cold outside to operating room. Finally, a ZCB00 (Allergan) was implanted.

CONCLUSIONS

The acute intraoperative clouding of trifocal lens(AT LISA tri 839mp) results from fluctuation of temperature should be noticed.

摘要

背景

有报道称,由于温度变化,几种亲水性丙烯酸人工晶状体(IOL)在术中出现晶状体混浊。此前报道的这种现象发生在寒冷国家和冬季。然而,尚无术中三焦点人工晶状体混浊的临床病例报道。我们报告一例在将AT LISA tri 839mp(卡尔·蔡司)植入眼内时,光学区同时发生急性混浊的病例。

病例介绍

一名79岁患有皮质和核性白内障的女性计划接受右眼飞秒技术辅助的超声乳化手术。选择植入由具有疏水表面特性的亲水性丙烯酸(25%)制成的三焦点人工晶状体(AT LISA tri 839mp)。当人工晶状体植入眼内时,它立即变得混浊。然后用从外部晶状体公司调来的另一枚AT LISA tri 839mp替换,观察到同样的现象。这两枚人工晶状体经历了从寒冷的外部环境到手术室相同的温度波动。最后植入了一枚ZCB00(爱尔康)人工晶状体。

结论

应注意到三焦点人工晶状体(AT LISA tri 839mp)术中急性混浊是由温度波动引起的。