在70厘米宽孔径临床3T MRI系统中使用质子磁共振波谱法对人伏隔核中谷氨酸进行定量分析的可靠性

Reliability of Glutamate Quantification in Human Nucleus Accumbens Using Proton Magnetic Resonance Spectroscopy at a 70-cm Wide-Bore Clinical 3T MRI System.

作者信息

Liu Xi-Long, Li Long, Li Jian-Neng, Rong Jia-Hui, Liu Bo, Hu Ze-Xuan

机构信息

Department of Radiology, Guangdong Provincial Corps Hospital of Chinese People's Armed Police Forces, Guangzhou Medical University, Guangzhou, China.

出版信息

Front Neurosci. 2017 Dec 5;11:686. doi: 10.3389/fnins.2017.00686. eCollection 2017.

Abstract

The human nucleus accumbens is a challenging region to study using proton magnetic resonance spectroscopy (H-MRS) on a 70-cm wide-bore clinical 3T MRI system. The aim of this study was to investigate the reliability for quantitative measurement of glutamate concentration in the nucleus accumbens using a 70-cm wide-bore clinical 3T MRI. H-MRS of the nucleus accumbens was acquired using the Point-Resolved Spectroscopic Sequence (PRESS) with echo time of 40 ms from 10 healthy volunteers (5 female; age range: 18-30 years) on two separate visits (a baseline, and 1-month time point). The Java-based Magnetic Resonance User Interface (jMRUI) software package was used to quantitatively measure the absolute metabolite concentrations. The test-retest reliability and reproducibility were assessed using intraclass correlations coefficients (ICC), and coefficients of variation (CV). Glutamate concentrations were similar across visits ( = 0.832). Reproducibility measures for all metabolites were good with CV ranging from 7.8 to 14.0%. The ICC values of all metabolites for the intra-class measures were excellent (ICC > 0.8), except that the reliability for Glx (glutamate + glutamine) was good (ICC = 0.768). Pearson correlations for all metabolites were all highly significant ( = 0.636-0.788, < 0.05). In conclusion, the short-echo-time PRESS can reliably obtain high quality glutamate spectrum from a ~3.4 cm voxel of the nucleus accumbens using a 70-cm wide-bore clinical 3T MRI.

摘要

在70厘米宽孔径的临床3T磁共振成像(MRI)系统上,使用质子磁共振波谱(H-MRS)研究人类伏隔核是一个具有挑战性的领域。本研究的目的是探讨使用70厘米宽孔径临床3T MRI定量测量伏隔核中谷氨酸浓度的可靠性。对10名健康志愿者(5名女性;年龄范围:18 - 30岁)在两次不同的就诊时(基线和1个月时间点),使用点分辨波谱序列(PRESS)、回波时间为40毫秒采集伏隔核的H-MRS。使用基于Java的磁共振用户界面(jMRUI)软件包定量测量绝对代谢物浓度。使用组内相关系数(ICC)和变异系数(CV)评估重测可靠性和可重复性。各次就诊间谷氨酸浓度相似( = 0.832)。所有代谢物的可重复性测量结果良好,CV范围为7.8%至14.0%。除了Glx(谷氨酸 + 谷氨酰胺)的可靠性良好(ICC = 0.768)外,所有代谢物的组内测量ICC值都非常出色(ICC > 0.8)。所有代谢物的Pearson相关性都非常显著( = 0.636 - 0.788, < 0.05)。总之,使用70厘米宽孔径临床3T MRI,短回波时间PRESS可以从伏隔核约3.4厘米的体素中可靠地获得高质量的谷氨酸谱。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e4d4/5723319/da241d7e821b/fnins-11-00686-g0001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验