Department of Chemistry, University of Warmia and Mazury, 10-957, Olsztyn, Poland.
Faculty of Science and Technology, Norwegian University of Life Sciences, Aas, Norway.
Environ Monit Assess. 2018 Feb 12;190(3):133. doi: 10.1007/s10661-018-6511-z.
The novel automatic flow system, direct-injection detector (DID) integrated with multi-pumping flow system (MPFS), dedicated for the photometric determination of orthophosphates in wastewater and freshwater samples is for the first time described. All reagents and the sample were injected simultaneously, in counter-current into the reaction-detection chamber by the system of specially selected for this purpose solenoid micro-pumps. The micro-pumps provided good precision and accuracy of the injected volumes. For the determination of orthophosphates, the molybdenum blue method was employed. The developed method can be used to detect orthophosphate in the range 0.1-12 mg L, with the repeatability (RSD) about 2.2% at 4 mg L and a very high injection throughput of 120 injections h. It was possible to achieve a very small consumption of reagents (10 μL of ammonium molybdate and 10 μL of ascorbic acid) and sample (20 μL). The volume of generated waste was only 440 μL per analysis. The method has been successfully applied, giving a good accuracy, to determination of orthophosphates in complex matrix samples: treated wastewater, lake water and reference sample of groundwater. The developed system is compact, small in both size and weight, requires 12 V in supply voltage, which are desirable for truly portable equipment used in routine analysis. The simplicity of the system should result in its greater long-time reliability comparing to other flow methods previously described.
首次描述了一种新颖的自动流动系统,即直接进样检测器(DID)与多泵送流动系统(MPFS)集成,专用于光度法测定废水和淡水样品中的正磷酸盐。所有试剂和样品都由专门为此目的选择的电磁阀微泵同时、逆流注入反应-检测室。微泵提供了良好的注入体积精度和准确性。对于正磷酸盐的测定,采用钼蓝法。所开发的方法可用于检测 0.1-12mg/L 范围内的正磷酸盐,在 4mg/L 时重复性(RSD)约为 2.2%,注射通量非常高,可达 120 次/小时。可以实现试剂(钼酸铵 10μL 和抗坏血酸 10μL)和样品(20μL)的非常小的消耗。每次分析生成的废物体积仅为 440μL。该方法已成功应用于复杂基质样品(处理后的废水、湖水和地下水参考样品)中正磷酸盐的测定,具有良好的准确性。所开发的系统紧凑、小巧,尺寸和重量都很小,仅需 12V 电源,非常适合用于常规分析的真正便携式设备。与以前描述的其他流动方法相比,系统的简单性应导致其更长的长期可靠性。