A theoretical analysis and finite element simulation of fixator-bone system stiffness on healing progression.

作者信息

Li Jianfeng, Zhao Xia, Hu Xiaojie, Tao Chunjing, Ji Run

机构信息

1 College of Mechanical Engineering and Applied Electronics Technology, Beijing University of Technology, Beijing, China.

2 National Research Center for Rehabilitation Technical Aids, Beijing, China.

出版信息

J Appl Biomater Funct Mater. 2018 Jul;16(3):115-125. doi: 10.1177/2280800017750357. Epub 2018 Mar 27.

Abstract

INTRODUCTION

The unilateral external fixator has become a quick and easy application for fracture stabilization of the extremities; the main value for evaluation of mechanical stability of the external fixator is stiffness. The stiffness property of the external fixator affects the local biomechanical environment of fractured bone.

METHODS

In this study, a theoretical model with changing Young's modulus of the callus is established by using the Castigliano's theory, investigating compression stiffness, torsional stiffness and bending stiffness of the fixator-bone system during the healing process. The effects of pin deviation angle on three stiffness methods are also investigated. In addition, finite element simulation is discussed regarding the stress distribution between the fixator and bone.

RESULTS

The results reveal the three stiffness evaluation methods are similar for the fixator-bone system. Finite element simulation shows that with increased healing time, the transmission of the load between the fixator and bone are different. In addition, the finite element analyses verify the conclusions obtained from the theoretical model.

CONCLUSIONS

This work helps orthopedic doctors to monitor the progression of fracture healing and determine the appropriate time for removal of a fixation device and provide important theoretical methodology.

摘要

引言

单侧外固定器已成为四肢骨折固定的一种快速简便的应用方式;评估外固定器机械稳定性的主要指标是刚度。外固定器的刚度特性会影响骨折部位的局部生物力学环境。

方法

在本研究中,利用卡氏定理建立了一个骨痂杨氏模量变化的理论模型,研究愈合过程中固定器 - 骨系统的压缩刚度、扭转刚度和弯曲刚度。还研究了针偏角对三种刚度测量方法的影响。此外,讨论了关于固定器与骨之间应力分布的有限元模拟。

结果

结果表明,固定器 - 骨系统的三种刚度评估方法相似。有限元模拟表明,随着愈合时间的增加,固定器与骨之间的载荷传递情况有所不同。此外,有限元分析验证了从理论模型得出的结论。

结论

本研究有助于骨科医生监测骨折愈合进程,确定拆除固定装置的合适时间,并提供重要的理论方法。