Suppr超能文献

一种使用二维 X 射线片进行 3D 膝关节假体规划和治疗评估的新技术:临床评估。

A novel technology for 3D knee prosthesis planning and treatment evaluation using 2D X-ray radiographs: a clinical evaluation.

机构信息

Institute for Surgical Technology and Biomechanics, University of Bern, Bern, Switzerland.

Clinic for Orthopedic, Sports Medicine and Rehabilitation, Krankenhaus Mrkisch Oderland GmbH, Wriezen, Germany.

出版信息

Int J Comput Assist Radiol Surg. 2018 Aug;13(8):1151-1158. doi: 10.1007/s11548-018-1789-4. Epub 2018 May 21.

Abstract

PURPOSE

To present a clinical validation of a novel technology called "3X" which allows for 3D prosthesis planning and treatment evaluation in total knee arthroplasty (TKA) using only 2D X-ray radiographs.

MATERIALS AND METHODS

After local institution review board approvals, 3X was evaluated on 43 cases (23 for preoperative planning and 20 for postoperative treatment evaluation). All the patients underwent CT scans according to a standard protocol. The results measured on the CT data were regarded as the ground truth. Additionally, two X-ray images were acquired for each affected leg and were used by 3X technology to derive patient-specific measurements of the leg. In total, we compared seven parameters for planning TKA and five parameters for postoperative prosthesis alignment.

RESULTS

Our experimental results demonstrated that the mean distances between the surface models reconstructed from 2D X-rays and the associated surface models obtained from 3D CT data were smaller than 1.5 mm. The average differences for all angular parameters were smaller than [Formula: see text]. In over 78% cases 3X technology derived the same femoral component size as the CT-based ground truth and this value went down to 70% when 3X technology was used to predict the size of tibial component.

CONCLUSION

3X is a technology that allows for true 3D preoperative planning and postoperative treatment evaluation based on 2D X-ray radiographs.

摘要

目的

介绍一种名为“3X”的新技术的临床验证,该技术仅使用 2D X 射线即可进行全膝关节置换术(TKA)的 3D 假体规划和治疗评估。

材料与方法

在获得当地机构审查委员会批准后,对 43 例患者(23 例用于术前规划,20 例用于术后治疗评估)进行了 3X 评估。所有患者均按照标准方案进行 CT 扫描。将 CT 数据上测量的结果视为基准。此外,为每条受影响的腿获取了两个 X 射线图像,并由 3X 技术用于得出患者腿部的特定测量值。总共比较了用于 TKA 规划的七个参数和用于术后假体对准的五个参数。

结果

我们的实验结果表明,从 2D X 射线重建的表面模型与从 3D CT 数据获得的相关表面模型之间的平均距离小于 1.5mm。所有角度参数的平均差异均小于[公式:见正文]。在超过 78%的情况下,3X 技术得出的股骨组件尺寸与基于 CT 的基准相同,而当 3X 技术用于预测胫骨组件的尺寸时,该值下降到 70%。

结论

3X 是一种技术,可基于 2D X 射线进行真正的 3D 术前规划和术后治疗评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验