Eur J Radiol. 2018 Jun;103:57-64. doi: 10.1016/j.ejrad.2018.04.011. Epub 2018 Apr 12.
The purpose of this study is to evaluate and compare 2-point (2pt), 3-point (3pt), and 6-point (6pt) Dixon magnetic resonance imaging (MRI) sequences with flexible echo times (TE) to measure proton density fat fraction (PDFF) within muscles. Two subject groups were recruited (G1: 23 young and healthy men, 31 ± 6 years; G2: 50 elderly men, sarcopenic, 77 ± 5 years). A 3-T MRI system was used to perform Dixon imaging on the left thigh. PDFF was measured with six Dixon prototype sequences: 2pt, 3pt, and 6pt sequences once with optimal TEs (in- and opposed-phase echo times), lower resolution, and higher bandwidth (optTE sequences) and once with higher image resolution (highRes sequences) and shortest possible TE, respectively. Intra-fascia PDFF content was determined. To evaluate the comparability among the sequences, Bland-Altman analysis was performed. The highRes 6pt Dixon sequences served as reference as a high correlation of this sequence to magnetic resonance spectroscopy has been shown before. The PDFF difference between the highRes 6pt Dixon sequence and the optTE 6pt, both 3pt, and the optTE 2pt was low (between 2.2% and 4.4%), however, not to the highRes 2pt Dixon sequence (33%). For the optTE sequences, difference decreased with the number of echoes used. In conclusion, for Dixon sequences with more than two echoes, the fat fraction measurement was reliable with arbitrary echo times, while for 2pt Dixon sequences, it was reliable with dedicated in- and opposed-phase echo timing.
本研究旨在评估和比较 2 点(2pt)、3 点(3pt)和 6 点(6pt)Dixon 磁共振成像(MRI)序列与灵活回波时间(TE),以测量肌肉内质子密度脂肪分数(PDFF)。招募了两组受试者(G1:23 名年轻健康男性,31±6 岁;G2:50 名老年男性,肌少症,77±5 岁)。使用 3T MRI 系统对左大腿进行 Dixon 成像。使用六个 Dixon 原型序列测量 PDFF:2pt、3pt 和 6pt 序列各一次,采用最佳 TE(同相和反相回波时间)、较低分辨率和较高带宽(optTE 序列),以及一次采用较高图像分辨率(高分辨率序列)和最短可能 TE。分别测定筋膜内 PDFF 含量。为了评估序列之间的可比性,进行了 Bland-Altman 分析。高分辨率 6pt Dixon 序列作为参考,因为之前已经证明该序列与磁共振波谱学具有高度相关性。高分辨率 6pt Dixon 序列与 optTE 6pt、3pt 和 optTE 2pt 之间的 PDFF 差异较低(介于 2.2%和 4.4%之间),但与高分辨率 2pt Dixon 序列(33%)不同。对于 optTE 序列,差异随使用的回波数量减少而减小。总之,对于具有两个以上回波的 Dixon 序列,任意回波时间的脂肪分数测量是可靠的,而对于 2pt Dixon 序列,专用的同相和反相回波定时是可靠的。