乳腺 MRI 在 BI-RADS 类别 4B 乳腺微钙化中的价值:基于 BI-RADS 第 5 版。

The value of breast MRI for BI-RADS category 4B mammographic microcalcification: based on the 5 edition of BI-RADS.

机构信息

Department of Radiology, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea.

Department of Radiology, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea.

出版信息

Clin Radiol. 2018 Aug;73(8):750-755. doi: 10.1016/j.crad.2018.04.014. Epub 2018 May 28.

Abstract

AIM

To investigate whether magnetic resonance imaging (MRI) can improve the positive predictive value (PPV) for Breast Imaging-Reporting and Data System (BI-RADS) category 4B mammographic microcalcification.

MATERIALS AND METHODS

One hundred and eight consecutive patients with BI-RADS category 4B microcalcification without mass on mammography underwent breast MRI and subsequent histopathological confirmation between January 2009 and December 2015. Mammography and MRI findings were reviewed retrospectively, and imaging features were analysed according to the 5 edition of BI-RADS. The PPV of each descriptor was analysed to identify subgroups in which PPV could be improved by the addition of MRI.

RESULTS

When the criteria of presence of enhancement on MRI was applied to category 4B microcalcification, PPV increased from 0.38 (41 of 108) to 0.82 (37 of 45) and reduced benign biopsy results by 88% (59 of 67). Four ductal carcinoma in situ lesions were missed. For amorphous microcalcification with regional or grouped distribution, MRI images increased PPV without missing malignancy.

CONCLUSION

Breast MRI has the potential to improve PPV for category 4B mammographic microcalcification by reducing false-positive findings. If amorphous microcalcification with regional or grouped distribution on mammography shows no enhancement on MRI, follow-up could be considered rather than immediate biopsy. In addition, breast MRI might have the potential to guide the best site to biopsy in category 4B microcalcification.

摘要

目的

探讨磁共振成像(MRI)是否能提高乳腺影像报告和数据系统(BI-RADS)4B 类乳腺钼靶微钙化的阳性预测值(PPV)。

材料与方法

2009 年 1 月至 2015 年 12 月,连续 108 例患者均为乳腺钼靶 BI-RADS 4B 类微钙化而无肿块,行乳腺 MRI 检查,并随后进行组织病理学检查。回顾性分析乳腺钼靶和 MRI 检查结果,根据 BI-RADS 第 5 版分析影像学特征。分析各描述符的 PPV,以确定通过 MRI 检查可改善哪些亚组的 PPV。

结果

当将 MRI 上存在强化的标准应用于 BI-RADS 4B 类微钙化时,PPV 从 0.38(108 例中的 41 例)增加到 0.82(45 例中的 37 例),并将良性活检结果减少了 88%(67 例中的 59 例)。有 4 例导管原位癌漏诊。对于区域性或成簇分布的无定形微钙化,MRI 图像增加了 PPV 而未遗漏恶性肿瘤。

结论

乳腺 MRI 通过减少假阳性发现,有可能提高 BI-RADS 4B 类乳腺钼靶微钙化的 PPV。如果乳腺钼靶上呈区域性或成簇分布的无定形微钙化在 MRI 上未见强化,可以考虑随访而不是立即进行活检。此外,乳腺 MRI 可能有潜力指导在 BI-RADS 4B 类微钙化中进行最佳活检部位。