Precision of Limb Length Measurement in Imageless Navigation THA with Modified Registration Technique in Semilateral Decubitus Position.

作者信息

Suksathien Yingong, Sueajui Jithayut

出版信息

J Med Assoc Thai. 2017 Jan;100(1):50-6.

Abstract

BACKGROUND

Limb length discrepancy (LLD) following total hip arthroplasty (THA) can be a significant problem and leading to patient dissatisfaction. Several researchers have reported that imageless navigation is a reliable technique and results in more precise cup placement compared to conventional freehand techniques, however, few studies have been reported about the accuracy of the femoral stem placement.

OBJECTIVE

(1) To evaluate the precision of an imageless navigation system in measuring the limb length change and (2) to evaluate LLD following imageless navigation THA with modified registration technique in semilateral decubitus position.

MATERIAL AND METHOD

The authors reviewed 66 cases receiving cementless THA with imageless navigation from September 2013 to December 2014. The radiographic limb length change measured from pre-operative and post-operative digital x-ray was compared with the intraoperative calculation by the navigation system. Postoperative LLD in unilateral cases and second operation of staged bilateral cases were also recorded

RESULTS

The mean radiographic limb length change measured on digital x-ray was 17.4 mm (5 to 29.3, SD 5.7). The mean limb length change calculated by navigation system was 16.8 mm (3 to 28,SD 5.9). The mean paired difference was 2.27 mm (-6 to 8, SD 0.9). This difference was significant (p = 0.01). There was significant correlation between LL change measured on digital x-ray and which were calculated by navigation system (r = 0.95, p<0.001). The navigation system had an accuracy of within 1 mm of the radiographic measurement in 7.6% of cases, within 2 mm in 39.4% of cases and within 5 mm in 93.9% of cases. The mean postoperative LLD was 2 mm (0 to 7, SD 1.9), 92% were within 5 mm.

CONCLUSION

The present study showed that the imageless navigation THA with modified registration technique in semilateral decubitus position offered a precise limb length measurement and the results encouraged us for clinical use to minimize LLD in THA.

摘要

背景

全髋关节置换术(THA)后肢体长度不等(LLD)可能是一个严重问题,并导致患者不满。一些研究人员报告称,与传统徒手技术相比,无影像导航是一种可靠的技术,能使髋臼放置更精确,然而,关于股骨柄放置准确性的研究报道较少。

目的

(1)评估无影像导航系统测量肢体长度变化的精度,(2)评估在半侧卧位采用改良注册技术的无影像导航THA后的LLD。

材料与方法

作者回顾了2013年9月至2014年12月期间接受无影像导航非骨水泥型THA的66例病例。将术前和术后数字X线片测量的影像学肢体长度变化与导航系统的术中计算结果进行比较。还记录了单侧病例的术后LLD以及分期双侧病例的二次手术情况。

结果

数字X线片测量的平均影像学肢体长度变化为17.4 mm(5至29.3,标准差5.7)。导航系统计算的平均肢体长度变化为16.8 mm(3至28,标准差5.9)。平均配对差异为2.27 mm(-6至8,标准差0.9)。这一差异具有统计学意义(p = 0.01)。数字X线片测量的LL变化与导航系统计算的变化之间存在显著相关性(r = 0.95,p<0.001)。导航系统在7.6%的病例中影像学测量精度在1 mm以内,39.4%的病例在2 mm以内,93.9%的病例在5 mm以内。平均术后LLD为2 mm(0至7,标准差1.9),92%在5 mm以内。

结论

本研究表明,在半侧卧位采用改良注册技术的无影像导航THA能提供精确的肢体长度测量,其结果鼓励我们在临床中使用,以尽量减少THA中的LLD。