Section of Cardiac Electrophysiology, University of Colorado, Aurora, Colorado.
Texas Cardiac Arrhythmia Institute, Austin, Texas.
JACC Clin Electrophysiol. 2018 Sep;4(9):1176-1185. doi: 10.1016/j.jacep.2018.06.021. Epub 2018 Aug 29.
This study sought to evaluate the efficacy and safety of using half-normal saline (HNS) as the cooling radiofrequency ablation (RFA) irrigant among patients who had failed prior, standard RFA.
Effective control of ventricular arrhythmias that arise from mid-myocardium may be refractory to standard RFA. Recent data suggest that delivering fluid with decreased ionic concentration during open-irrigated RFA can produce deeper RFA lesions.
A 12-center prospective analysis was performed of all ablations using HNS for the treatment of ventricular arrhythmias (premature ventricular complex [PVC]/ventricular tachycardia [VT]) refractory to standard ablation with normal saline irrigant.
HNS RFA was used clinically to target 94 PVC/VTs refractory to standard ablation. Acute success was achieved in 78 of 94 (83%), with longer-term success occurring in 78 subjects after a mean follow-up of 6.1 ± 6.7 months (range, 3.0 to 25.2 months). Steam pops were observed among 12 (12.6%) patients. There were no significant changes in electrolytes measured before and after the use of HNS, and there were no complications related to HNS use.
The use of HNS instead of normal saline irrigant during high-power delivery targeting deep myocardial substrate is safe and effective. PVC/VT sources previously unaffected by standard ablation may be successfully ablated with improved efficiency of radiofrequency delivery using HNS.
本研究旨在评估在先前标准射频消融 (RFA) 失败的患者中使用半生理盐水 (HNS) 作为冷却 RFA 灌流液的疗效和安全性。
有效控制起源于心中层的室性心律失常可能对标准 RFA 有抗性。最近的数据表明,在开放式灌流 RFA 期间输送离子浓度降低的液体可以产生更深的 RFA 病变。
对所有使用 HNS 治疗因生理盐水灌流而导致的室性心律失常(室性早搏 [PVC]/室性心动过速 [VT])的消融进行了 12 中心前瞻性分析,这些心律失常对标准消融有抗性。
HNS RFA 临床上用于靶向 94 例 PVC/VT,这些病例对标准消融有抗性。94 例中的 78 例(83%)急性成功,78 例在平均随访 6.1±6.7 个月(范围 3.0 至 25.2 个月)后获得长期成功。12 例(12.6%)患者观察到蒸汽声。使用 HNS 前后电解质测量无显著变化,且无与 HNS 使用相关的并发症。
在针对深层心肌底物的高功率输送中使用 HNS 代替生理盐水灌流是安全有效的。使用 HNS 可提高射频输送效率,以前不受标准消融影响的 PVC/VT 源可能被成功消融。