To assess the transferability of the magnetic resonance imaging (MRI) multislice multiecho T2* technique for pancreatic iron overload assessment.
Multiecho T2* sequences were installed on ten 1.5-T MRI scanners of the three main vendors. Five healthy subjects (n = 50) were scanned at each site. Five patients with thalassemia (n = 45) were scanned locally at each site and were rescanned at the reference site within 1 month. T2* images were analyzed using a previously validated software and the global pancreatic T2* value was calculated as the mean of T2* values over the head, body, and tail.
T2* values of healthy subjects were above 26 ms and showed inter-site homogeneity. The T2* values measured in the MRI sites were comparable to the correspondent values observed in the reference site (12.02 ± 10.20 ms vs 11.98 ± 10.47 ms; p = 0.808), and the correlation coefficient was 0.978 (p < 0.0001). Coefficients of variation (CoVs) ranged from 4.22 to 9.77%, and the CoV for all the T2* values independently from the sites was 8.55%. The intraclass correlation coefficient (ICC) for each MRI site was always excellent and the global ICC was 0.995, independently from the sites. The mean absolute difference in patients with pancreatic iron (n = 39) was -0.15 ± 1.38 ms.
The gradient-echo T2* MRI technique is an accurate and reproducible means for the quantification of pancreatic iron and may be transferred among MRI scanners by different vendors in several centers.
• The gradient-echo T2* MRI technique is an accurate and reproducible means for the quantification of pancreatic iron. • The gradient-echo T2* MRI technique for the quantification of pancreatic iron may be transferred among MRI scanners by different vendors in several centers. • Pancreatic iron might serve as an early predictor of cardiac siderosis and is the strongest overall predictor of glucose dysregulation.
评估磁共振成像(MRI)多层多回波 T2*技术用于胰腺铁过载评估的可转移性。
在三个主要供应商的十台 1.5-T MRI 扫描仪上安装多回波 T2序列。每个站点扫描 5 名健康受试者(n=50)。每个站点对 5 名地中海贫血患者(n=45)进行局部扫描,并在 1 个月内在参考站点进行重新扫描。使用先前验证的软件分析 T2图像,并计算头部、体部和尾部 T2值的平均值作为胰腺的整体 T2值。
健康受试者的 T2值高于 26 ms,表现出站点间的同质性。MRI 站点测量的 T2值与参考站点观察到的对应值相似(12.02±10.20 ms 与 11.98±10.47 ms;p=0.808),相关系数为 0.978(p<0.0001)。变异系数(CoV)范围为 4.22%至 9.77%,所有 T2*值的 CoV 独立于站点为 8.55%。每个 MRI 站点的组内相关系数(ICC)始终为优秀,全局 ICC 为 0.995,独立于站点。胰腺铁(n=39)患者的平均绝对差值为-0.15±1.38 ms。
梯度回波 T2* MRI 技术是一种准确且可重复的胰腺铁定量方法,可在不同供应商的多个中心的 MRI 扫描仪之间转移。
• 梯度回波 T2* MRI 技术是一种准确且可重复的胰腺铁定量方法。
• 梯度回波 T2* MRI 技术可在不同供应商的多个中心的 MRI 扫描仪之间转移,用于胰腺铁的定量。
• 胰腺铁可能作为心脏铁过载的早期预测指标,也是葡萄糖调节紊乱的最强总体预测指标。