Couette shearing device for the investigation of shear-induced damage of the primary hemostasis by left ventricular assist devices.

作者信息

Lommel Michael Achim, Goubergrits Leonid, Affeld Klaus, Kertzscher Ulrich

机构信息

Biofluid Mechanics Laboratory, Institute for Imaging Science and Computational Modelling in Cardiovascular Medicine, Charité-Universitätsmedizin Berlin, Berlin, Germany.

出版信息

Int J Artif Organs. 2019 Mar;42(3):143-150. doi: 10.1177/0391398818802950. Epub 2018 Oct 22.

Abstract

INTRODUCTION

: Continuous-flow left ventricular assist devices have evolved from short-time therapy into permanent or so-called destination therapy. One complication in long-term usage is bleeding, which is presumably attributed to shear-induced interference of left ventricular assist devices with the coagulation system.

METHODS

: The influence of dynamic shear stresses on primary hemostasis by single or multiple passes through left ventricular assist devices was investigated. A novel Couette-type shearing device, especially fitted to simulate left ventricular assist devices with highly dynamic and repetitive stresses, was developed. To evaluate the clotting ability of the blood and thus the bleeding tendency, the closure time of the platelet function analyzer (PFA-100, Dade Behring, Marburg, Germany) was used. The relationship of the PFA-100 closure time was fitted to measurement points with shear stress and exposure time as parameters.

RESULTS

: 76 samples of human blood collected from four different healthy donors in sodium-citrate anticoagulant solution were tested, including 20 control samples. A damage model according to the power law approach could be developed. A linear correlation of shear stress and exposure time to the PFA-100 closure time could be determined. In addition, a model was developed to calculate the increase in the PFA closure time on the basis of shear stress over time curves.

DISCUSSION

: With the shearing device, half-sine-wave-shaped shear stress patterns relevant to rotary blood pumps can be achieved with very good repeatability. The proposed damage model could be used to compare and optimize left ventricular assist devices under development. The tests showed a significant decrease in coagulability after only a few repetitions.

摘要

引言

连续流左心室辅助装置已从短期治疗发展为永久性或所谓的目标治疗。长期使用中的一个并发症是出血,这可能归因于左心室辅助装置对凝血系统的剪切诱导干扰。

方法

研究了单次或多次通过左心室辅助装置时动态剪切应力对初级止血的影响。开发了一种新型库埃特型剪切装置,特别适合模拟具有高度动态和重复应力的左心室辅助装置。为了评估血液的凝血能力以及出血倾向,使用了血小板功能分析仪(PFA-100,德国马尔堡达德拜林公司)的关闭时间。以剪切应力和暴露时间为参数,将PFA-100关闭时间与测量点进行拟合。

结果

测试了从四名不同健康供体采集的76份柠檬酸钠抗凝溶液中的人体血液样本,包括20份对照样本。可以建立一个基于幂律方法的损伤模型。可以确定剪切应力和暴露时间与PFA-100关闭时间之间的线性相关性。此外,还开发了一个模型,用于根据剪切应力随时间的曲线计算PFA关闭时间的增加。

讨论

使用该剪切装置,可以实现与旋转血泵相关的半正弦波形状的剪切应力模式,且具有非常好的重复性。所提出的损伤模型可用于比较和优化正在开发的左心室辅助装置。测试表明,仅经过几次重复后,凝血能力就显著下降。