Abstract

OBJECTIVES

The aim of this study was to test the accuracy of different methods of resting energy expenditure (REE) data analysis using indirect calorimetry (IC) during traditional (30 min) and abbreviated (10 min) protocols.

METHODS

Fifteen women and 15 men (21-34 y of age) completed two consecutive 30-min IC measurements. Body composition was measured using dual-energy x-ray absorptiometry. The reference method for REE analysis was 5 min in steady state (SS) during 30 min (first 5 min discarded). REE measurements were randomized to define a reference or testing method. An interval method was defined using 25, 20, and 15 min (with first 5, 10, and 15 min discarded, respectively), during 30 min, and 5 min (first 5 min discarded) during 10-min intervals. The SS method was defined using 5 min in SS (first 5 min discarded) during 30 min, 5, 4, and 3 min in SS during 10-min (first 5 min discarded) intervals.

RESULTS

Interval methods during 30 min and SS and interval methods during 10 min demonstrated large bias with significantly high REEs compared to the reference method (78.8-109.0 kcal/d, all P < 0.001). Testing methods demonstrated large upper limits of agreement between 225.2 and 322.8 kcal/d. No mean differences (P > 0.05), small bias (14.3 kcal/d), and narrow limits of agreement (-125.8 to 154.4 kcal/d) were observed between 5-min SS during 30 min and the reference method.

CONCLUSIONS

All interval methods and SS methods during 10 min overestimated REE. We recommend using 5-min SS during 30 min. The measurement may be repeated until all participants achieve SS.

摘要

目的

本研究旨在测试间接测热法(IC)在传统(30 分钟)和简化(10 分钟)方案中使用不同的静息能量消耗(REE)数据分析方法的准确性。

方法

15 名女性和 15 名男性(21-34 岁)连续完成了两次 30 分钟的 IC 测量。使用双能 X 射线吸收法测量身体成分。REE 分析的参考方法是在 30 分钟内稳定状态(SS)持续 5 分钟(先丢弃前 5 分钟)。REE 测量随机分配以定义参考或测试方法。间隔法定义为在 30 分钟内使用 25、20 和 15 分钟(分别丢弃前 5、10 和 15 分钟),以及在 10 分钟间隔内使用 5 分钟(先丢弃前 5 分钟)。SS 法在 30 分钟内使用 SS 持续 5 分钟(先丢弃前 5 分钟),在 10 分钟内使用 SS 持续 5、4 和 3 分钟(先丢弃前 5 分钟)。

结果

30 分钟间隔法和 SS 以及 10 分钟间隔法和 SS 与参考方法相比,REE 显著升高,具有较大的偏差(78.8-109.0 kcal/d,均 P <0.001)。测试方法之间的上界差异较大,为 225.2 至 322.8 kcal/d。在 30 分钟期间,5 分钟 SS 与参考方法之间没有观察到平均差异(P > 0.05)、小偏差(14.3 kcal/d)和窄的一致性界限(-125.8 至 154.4 kcal/d)。

结论

所有间隔法和 10 分钟 SS 方法均高估了 REE。我们建议在 30 分钟内使用 5 分钟 SS。可以重复测量,直到所有参与者达到 SS。