The aim of this study was to investigate the impact on metal artifacts and dose efficiency of using a tin filter in combination with high-energy threshold (TH) images of a photon-counting detector (PCD) computed tomography (CT) system.
A 3D-printed spine with pedicle screws was scanned on a PCD-CT system with and without tin filtration. Image noise and severity of artifacts were measured for low-energy threshold (TL) and TH images. In a prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant study, 20 patients having a clinical energy-integrating detector (EID) CT were scanned on a PCD-CT system using tin filtration. Images were reviewed by 3 radiologists to evaluate visualization of anatomic structures, diagnostic confidence, and image preference. Artifact severity and image noise were measured. Wilcoxon signed rank was used to test differences between PCD-CT TH and EID-CT images.
Phantom TH images with tin filtration reduced metal artifacts and had comparable noise (32 HU) to TL images (29 HU) acquired without tin filtration. Visualization scores for the cortex, trabeculae, and implant-trabecular interface from PCD-CT TH images (4.4 ± 0.9, 4.4 ± 1.0, and 4.4 ± 1.0) were significantly higher (P < 0.0001) than EID-CT images (3.3 ± 1.3, 3.3 ± 1.2, and 3.3 ± 1.6). A strong preference was shown for PCD-CT TH images due to improved diagnostic confidence and decreased artifact severity. Noise in PCD-CT TH images (93 ± 41 HU) was significantly lower than that in EID-CT images (133 ± 92 HU, P < 0.05).
Threshold high images acquired with tin filtration on PCD-CT demonstrated a substantial decrease in metal artifacts and an increase in dose efficiency compared with EID-CT.
本研究旨在探讨在光子计数探测器(PCD)计算机断层扫描(CT)系统中使用锡滤过片结合高能阈值(TH)图像对金属伪影和剂量效率的影响。
使用 PCD-CT 系统对带有椎弓根螺钉的 3D 打印脊柱模型进行扫描,分别在使用和不使用锡滤过片的情况下进行。对低能阈值(TL)和 TH 图像进行图像噪声和伪影严重程度的测量。在一项前瞻性的、经机构审查委员会批准的、符合《健康保险流通与责任法案》的研究中,对 20 例接受临床能量积分探测器(EID)CT 检查的患者在 PCD-CT 系统上使用锡滤过片进行扫描。由 3 名放射科医生对图像进行评估,以评价解剖结构的可视性、诊断信心和图像偏好。测量了伪影严重程度和图像噪声。采用 Wilcoxon 符号秩检验比较 PCD-CT TH 和 EID-CT 图像之间的差异。
在 PCD-CT TH 图像中使用锡滤过片降低了金属伪影,其噪声(32 HU)与未使用锡滤过片的 TL 图像(29 HU)相当。PCD-CT TH 图像的皮质、小梁和植入物-小梁界面的可视评分(4.4 ± 0.9、4.4 ± 1.0 和 4.4 ± 1.0)明显高于 EID-CT 图像(3.3 ± 1.3、3.3 ± 1.2 和 3.3 ± 1.6)(P < 0.0001)。由于诊断信心提高和伪影严重程度降低,强烈倾向于使用 PCD-CT TH 图像。PCD-CT TH 图像的噪声(93 ± 41 HU)明显低于 EID-CT 图像(133 ± 92 HU,P < 0.05)。
与 EID-CT 相比,在 PCD-CT 上使用锡滤过片获取的 TH 图像在降低金属伪影和提高剂量效率方面具有显著优势。