Suppr超能文献

远红外光谱范围内光学量的背景校正测量。

Background corrected measurements of optical quantities in the far-infrared spectral range.

作者信息

Kehrt M, Monte C, Steiger A, Hollandt J

出版信息

Opt Express. 2018 Dec 24;26(26):34002-34006. doi: 10.1364/OE.26.034002.

Abstract

By definition, optical quantities transmittance and reflectance can basically be determined as the ratio of two flux measurements. One measurement is performed with, and the other without, the sample under test in the optical path. However, at longer wavelengths the temperature radiation of the sample itself as well as of the applied spectrometer and detector increasingly contribute to the detected radiation budget. This leads to growing systematic errors in the determination of the transmittance and reflectance of samples with Fourier transform infrared spectrometers at longer wavelengths. We present an effective method to overcome this problem by measuring a sequence of four measurements at two different flux levels. Results obtained with this method are compared to the basic ratio method over a spectral range from 200 cm to 30 cm (0.9 THz to 6 THz).

摘要

根据定义,光学量透过率和反射率基本上可以确定为两个通量测量值的比值。一次测量是在光路中有被测样品的情况下进行的,另一次测量是在光路中没有被测样品的情况下进行的。然而,在较长波长下,样品本身以及所使用的光谱仪和探测器的热辐射对检测到的辐射量的贡献越来越大。这导致在使用傅里叶变换红外光谱仪测定较长波长样品的透过率和反射率时,系统误差不断增大。我们提出了一种有效的方法来克服这个问题,即在两个不同的通量水平下进行一系列四次测量。在200厘米至30厘米(0.9太赫兹至6太赫兹)的光谱范围内,将用这种方法获得的结果与基本比值法的结果进行了比较。