Clinically evaluate intraocular pressure (IOP) measurements taken with a Goldmann applanation tonometer (GAT) prism and a modified surface Goldmann prism examining measurement differences correlated to central corneal thickness (CCT) and corneal hysteresis (CH) values.
Prospective, open-label, randomised, controlled, multicentre reference device accuracy analysis.
A GAT and a modified surface GAT prism measured IOP on 243 unique eyes. The study design and methodology complied with International Standard Organization (ISO) tonometer evaluation guidelines, except the inclusion of thin (<500 µm) and thick (>600 µm) corneas. All eyes were randomised to IOP measurement by one of five standard Goldmann prisms and five modified prisms. Pressures were measured by six investigators, two times with each prism for a total of 1936 IOP measurements. Analysis included a multiple linear regression including CCT and CH correlation.
The difference in IOP measurements of the standard and modified Goldmann prisms correlated well to CCT particularly in thin (<500 µm) and thick (>600 µm) corneas (R=0.404, p=0.007). Corneal hysteresis (CH) also significantly correlated to the difference in prism measurements (R=0.125, p=0.039). There was no significant overall mean IOP bias between the two prisms (+0.43 mm Hg in modified, p=0.19).
The paired IOP measurement difference between GAT and a modified surface Goldmann replacement prism indicated a statistically significant correlation to CCT and CH. A simple modified replacement prism for any Goldmann-type tonometer may significantly improve IOP measurement accuracy by minimising corneal biomechanical errors associated with CCT and CH.
NCT02990169 and NCT02989909.
临床评估使用 Goldmann 压平眼压计(GAT)棱镜和改良表面 Goldmann 棱镜进行眼压(IOP)测量的结果,比较与中央角膜厚度(CCT)和角膜滞后量(CH)值相关的测量差异。
前瞻性、开放标签、随机、对照、多中心参考设备准确性分析。
GAT 和改良表面 GAT 棱镜对 243 只独特的眼睛进行了 IOP 测量。研究设计和方法符合国际标准化组织(ISO)眼压计评估指南,除了包括薄(<500 µm)和厚(>600 µm)角膜。所有眼睛随机分为一组,每组用 5 个标准 Goldmann 棱镜和 5 个改良棱镜进行 IOP 测量。由 6 名研究者进行压力测量,每个棱镜测量两次,总共进行了 1936 次 IOP 测量。分析包括包括 CCT 和 CH 相关性的多元线性回归。
标准和改良 Goldmann 棱镜的 IOP 测量差值与 CCT 相关性较好,尤其是在薄(<500 µm)和厚(>600 µm)角膜中(R=0.404,p=0.007)。角膜滞后量(CH)也与棱镜测量差值显著相关(R=0.125,p=0.039)。两种棱镜之间的平均眼压无显著差异(改良棱镜组为+0.43 mm Hg,p=0.19)。
GAT 和改良表面 Goldmann 替换棱镜的配对 IOP 测量差值与 CCT 和 CH 呈统计学显著相关。对于任何 Goldmann 眼压计,一个简单的改良替换棱镜可能通过最大限度地减少与 CCT 和 CH 相关的角膜生物力学误差,显著提高眼压测量的准确性。
NCT02990169 和 NCT02989909。