J Cataract Refract Surg. 2019 Jul;45(7):985-991. doi: 10.1016/j.jcrs.2019.02.005. Epub 2019 Apr 24.
To compare the difference and agreement of corneal higher-order aberrations (HOAs) in keratoconic eyes using Scheimpflug and dual Scheimpflug-Placido imaging systems.
Emory University, Atlanta, Georgia, and the University of Southern California, Los Angeles, California, USA.
Retrospective between-devices reliability and agreement study.
Patients diagnosed with keratoconus were evaluated sequentially by Scheimpflug and dual Scheimpflug-Placido devices. Differences, correlations, and agreement between values for total root mean square (RMS), trefoil, coma, and spherical aberration were analyzed, and Bland-Altman plots were generated.
Fifty eyes from 50 patients (31 men, 19 women) were evaluated. Trefoil at 30 degrees, spherical aberration, and total RMS were significantly different between groups (P < .05), whereas trefoil at 0 degrees and total coma values were not statistically different. There was a weak positive correlation between devices for trefoil at 0 degrees (r = 0.228), and a moderate positive correlation for trefoil at 30 degrees (r = 0.473), horizontal coma (r = 0.430), and for total corneal RMS (r = 0.637). Vertical coma (r = 0.816) and spherical aberration (r = 0.874) showed a strong positive correlation. The 95% limits of agreement (LoA) for absolute values were 1.963 μm for trefoil at 30 degrees, 2.449 μm for trefoil at 0 degrees, 3.530 μm for horizontal coma, 2.145 μm for vertical coma, 1.242 μm for spherical aberration, and 10.527 μm for RMS.
Significant differences were found between measurements of corneal HOAs generated by Scheimpflug and dual Scheimpflug-Placido devices in patients with keratoconus, with generally limited correlations and wide LoA. HOAs measurements from these devices should not be considered equivalent.
比较使用 Scheimpflug 系统和双 Scheimpflug-Placido 成像系统测量圆锥角膜眼角膜高阶像差(HOA)的差异和一致性。
美国佐治亚州亚特兰大市埃默里大学和加利福尼亚州洛杉矶市南加州大学。
回顾性设备间可靠性和一致性研究。
对被诊断为圆锥角膜的患者依次使用 Scheimpflug 和双 Scheimpflug-Placido 设备进行评估。分析总均方根(RMS)、三叶形、彗差和球差的差异、相关性和一致性,并生成 Bland-Altman 图。
对 50 例(31 例男性,19 例女性)50 只眼进行了评估。30°三叶形、球差和总 RMS 组间差异有统计学意义(P<0.05),而 0°三叶形和总彗差值无统计学差异。两组设备间 0°三叶形(r=0.228)和 30°三叶形(r=0.473)、水平彗差(r=0.430)和总角膜 RMS(r=0.637)相关性较弱;垂直彗差(r=0.816)和球差(r=0.874)相关性较强。绝对值的 95%一致性界限(LoA)分别为 30°三叶形 1.963μm、0°三叶形 2.449μm、水平彗差 3.530μm、垂直彗差 2.145μm、球差 1.242μm 和 RMS 10.527μm。
圆锥角膜患者的 Scheimpflug 和双 Scheimpflug-Placido 设备测量角膜 HOA 存在显著差异,通常相关性有限,一致性界限较宽。这些设备的 HOA 测量值不应被认为是等效的。