J Virol Methods. 2019 Aug;270:137-145. doi: 10.1016/j.jviromet.2019.05.007. Epub 2019 May 20.
Real-time PCR are often used for the diagnosis and monitoring of Cytomegalovirus (CMV) and Epstein-Barr virus (EBV) infections in susceptible populations. In this context, we evaluated the analytical performances of the Abbott RealTime CMV/EBV maxCycle protocol automated on the m2000 platform (Abbott). It was compared to our routinely-used procedure consisting of a NucleoMag® DNA extraction automated on a STARlet platform followed by manually processed CMV and EBV quantitative real-time PCR (Diagenode). In this study, we showed that both EBV assays exhibited a similar sensitivity but with a better precision for the EBV Abbott RealTime assay. For the CMV performances, the Abbott assay was more sensitive and more precise than our routine method. The use of WHO International Standards also indicated a slight underestimation of the viral loads (-0.25 log IU/mL and -0.21 log IU/mL for CMV and EBV assays respectively) while these were rather overestimated with the Starlet/Diagenode method (0.48 log IU/mL and 0.19 log IU/mL for CMV and EBV assays respectively). These trends were confirmed using relevant whole-blood clinical samples and external quality controls. The workflows were also compared and we highlighted a significant technician hands-on time reduction (-63%) using the Abbott CMV/EBV maxCycle automated protocol.
实时 PCR 常用于易感染人群中巨细胞病毒 (CMV) 和 Epstein-Barr 病毒 (EBV) 感染的诊断和监测。在这种情况下,我们评估了 Abbott RealTime CMV/EBV maxCycle 方案在 m2000 平台上(Abbott)的分析性能。该方案与我们常用的程序进行了比较,该程序包括在 STARlet 平台上自动进行的 NucleoMag® DNA 提取,然后对 CMV 和 EBV 进行手动定量实时 PCR(Diagenode)处理。在这项研究中,我们表明,两种 EBV 检测方法均表现出相似的灵敏度,但 EBV Abbott RealTime 检测法的精密度更好。对于 CMV 性能,Abbott 检测法比我们的常规方法更灵敏、更精确。使用世界卫生组织国际标准也表明,病毒载量存在轻微低估(CMV 和 EBV 检测分别为-0.25 log IU/mL 和-0.21 log IU/mL),而 Starlet/Diagenode 方法则高估了这些(CMV 和 EBV 检测分别为 0.48 log IU/mL 和 0.19 log IU/mL)。使用相关的全血临床样本和外部质量控制对这些趋势进行了验证。还比较了工作流程,并强调使用 Abbott CMV/EBV maxCycle 自动化方案可显著减少技术员的手动操作时间(减少 63%)。