Suppr超能文献

分段式食管自膨式金属支架力学性能评估:需要创新测试方法。

Evaluation of Mechanical Properties of Segmented Esophageal Self-Expandable Metal Stents: Innovative Test Methods Are Needed.

作者信息

Bürger Martin, Herbold Till, Lange Sven, Berlth Felix, Plum Patrick Sven, Schramm Christoph, Kleinert Robert, Goeser Tobias, Bruns Christiane Josephine, Chon Seung-Hun

机构信息

Department of Gastroenterology and Hepatology, University Hospital Cologne, Cologne, Germany.

Department of General, Visceral and Transplantation Surgery, RWTH Aachen, Aachen, Germany.

出版信息

J Laparoendosc Adv Surg Tech A. 2019 Sep;29(9):1168-1173. doi: 10.1089/lap.2019.0173. Epub 2019 Jun 4.

Abstract

Self-expanding metal stents (SEMSs) in different geometric shapes are well established treatment options in diseases of the esophagus. Mechanical properties and stent design may have an impact on patient comfort, migration rate, and removability. In this study, we evaluated mechanical properties of three segmented SEMSs (segSEMSs) for the esophagus with regard to distinct stent sections. Radial forces were measured using a testing method distinguishing between circumferential radial and local radial force. The center parts of the segSEMSs were measured for circumferential radial forces without being affected by the flared ends. Axial forces were measured at 20° bending. Circumferential radial force measurements over the full stent length showed substantial differences against measurements of the center parts of the stents as the flared ends falsify test results by up to 53%. Although circumferential radial forces of the center parts were about the same (<10% variances) for all segSEMSs, local radial forces showed considerable differences of up to 26%. One segSEMS showed high axial forces, whereas the other two only needed half of the force (up to 53%) to be bent to 20°. Flared ends of segSEMSs have a substantial impact on radial force measurements and therefore alter test results, confirmed by our separated center part test of segSEMSs. Our innovative setup whereby we compressed the stent in an asymmetric manner (local radial force) and evaluated sections of stents separately, indeed revealed differences to circumferential measurements, leading to a more in-depth knowledge of stent characteristics.

摘要

不同几何形状的自膨式金属支架(SEMS)是治疗食管疾病的成熟选择。机械性能和支架设计可能会影响患者舒适度、迁移率和可移除性。在本研究中,我们针对食管的三种分段式SEMS(segSEMS)的不同支架部分评估了其机械性能。使用一种区分周向径向力和局部径向力的测试方法来测量径向力。在不受喇叭口端影响的情况下测量segSEMS的中心部分的周向径向力。在20°弯曲时测量轴向力。对整个支架长度进行的周向径向力测量显示,与支架中心部分的测量结果存在显著差异,因为喇叭口端会使测试结果的偏差高达53%。尽管所有segSEMS的中心部分的周向径向力大致相同(方差<10%),但局部径向力显示出高达26%的显著差异。一种segSEMS显示出较高的轴向力,而其他两种只需一半的力(高达53%)就能弯曲到20°。segSEMS的喇叭口端对径向力测量有重大影响,从而改变测试结果,这一点已通过我们对segSEMS的单独中心部分测试得到证实。我们创新的设置方式,即不对称压缩支架(局部径向力)并分别评估支架各部分,确实揭示了与周向测量结果的差异,从而更深入地了解了支架特性。