IEEE Trans Biomed Circuits Syst. 2019 Oct;13(5):927-937. doi: 10.1109/TBCAS.2019.2925713. Epub 2019 Jul 3.
A highly integrated, wearable electrical impedance tomography (EIT) belt for neonatal thorax vital multiple sign monitoring is presented. The belt has 16 active electrodes. Each electrode has an application-specific integrated circuit (ASIC) connected to it. The ASIC contains a fully differential current driver, a high-performance instrumentation amplifier, a digital controller, and multiplexors. The belt features a new active electrode architecture that allows programmable flexible electrode current drive and voltage sense patterns under simple digital control. It provides intimate connections to the electrodes for the current drive and to the IA for direct differential voltage measurement, providing superior common-mode rejection ratio. The ASIC was designed in a CMOS 0.35-μm high-voltage technology. The high-specification EIT belt has an image frame rate of 122 fps, a wide operating bandwidth of 1 MHz, and multi-frequency operation. It measures impedance with 98% accuracy and has less than 0.5 Ω and 1° variation across all possible channels. The image results confirmed the advantage of the new active electrode architecture and the benefit of wideband, multi-frequency EIT operation. The system successfully captured high-quality lung-respiration EIT images, breathing cycle, and heart rate. It can also provide boundary-shape information by using an array of MEMS sensors interfaced to the ASICs.
一种高度集成的、可穿戴的新生儿胸部生命多体征监测用电阻抗断层成像(EIT)腰带被提出。该腰带拥有 16 个主动电极。每个电极都连接有一个专用集成电路(ASIC)。ASIC 中包含一个全差分电流驱动器、一个高性能仪表放大器、一个数字控制器和多路复用器。该腰带采用了一种新的主动电极架构,可在简单的数字控制下实现可编程的灵活电极电流驱动和电压感应模式。它为电流驱动提供了与电极的紧密连接,为 IA 提供了直接差分电压测量的连接,从而提供了卓越的共模抑制比。ASIC 采用了 0.35-μm 高压工艺的 CMOS 技术进行设计。这款高规格的 EIT 腰带具有 122 fps 的图像帧率、1 MHz 的宽工作带宽和多频操作。它的阻抗测量精度达到 98%,在所有可能的通道中,其变化小于 0.5 Ω 和 1°。图像结果证实了新型主动电极架构的优势,以及宽带、多频 EIT 操作的好处。该系统成功地捕捉到了高质量的肺部呼吸 EIT 图像、呼吸周期和心率。它还可以通过使用连接到 ASIC 的一组 MEMS 传感器来提供边界形状信息。