PLoS One. 2019 Jul 31;14(7):e0220521. doi: 10.1371/journal.pone.0220521. eCollection 2019.
The reliability of MyotonPRO that can monitor the mechanical properties of tissues is still unclear. This study aimed to analyze the within-day inter-operator and between-day intra-operator reliability of MyotonPRO for assessing tone and stiffness of quadriceps femoris and patellar tendon at different knee angles. The tone and stiffness of healthy participants (15 males and 15 females, aged 24.7±1.6 years) in the supine and resting position were measured using the MyotonPRO device. The measurements were quantified at 0°, 30°, 60°, and 90° of knee flexion. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated and a Bland-Altman analysis was conducted to estimate reliability. The results indicated excellent inter-operator reliability (ICC > 0.78) and good to excellent intra-operator reliability (ICC > 0.41). The inter-operator SEM measurements ranged between 0.1-0.9 Hz and 3.8-37.9 N/m, and intra-operator SEM ranged between 0.5-1.3 Hz and 7.9-52.0 N/m. The inter-operator MDC ranged between 0.3-2.5 Hz and 10.5-105.1 N/m, and intra-operator SEM ranged between 1.1-3.3 Hz and 21.9-144.1 N/m. The agreement of inter-operator was better than that of intra-operator. The study concluded that MyotonPRO is a reliable device to detect the tone and stiffness of quadriceps femoris and patellar tendon.
MyotonPRO 监测组织力学特性的可靠性尚不清楚。本研究旨在分析 MyotonPRO 评估不同膝关节角度下股四头肌和髌腱紧张度和僵硬度的日内操作者间和日间操作者内的可靠性。使用 MyotonPRO 设备测量仰卧位和休息位健康参与者(15 名男性和 15 名女性,年龄 24.7±1.6 岁)的紧张度和僵硬度。在膝关节屈曲 0°、30°、60°和 90°时对测量值进行量化。计算了组内相关系数(ICC)、测量标准误差(SEM)和最小可检测变化(MDC),并进行了 Bland-Altman 分析以评估可靠性。结果表明,操作者间具有极好的可靠性(ICC>0.78),操作者内具有良好至极好的可靠性(ICC>0.41)。操作者间 SEM 测量值范围在 0.1-0.9 Hz 和 3.8-37.9 N/m 之间,操作者内 SEM 范围在 0.5-1.3 Hz 和 7.9-52.0 N/m 之间。操作者间 MDC 范围在 0.3-2.5 Hz 和 10.5-105.1 N/m 之间,操作者内 SEM 范围在 1.1-3.3 Hz 和 21.9-144.1 N/m 之间。操作者间的一致性优于操作者内的一致性。该研究得出结论,MyotonPRO 是一种可靠的检测股四头肌和髌腱紧张度和僵硬度的设备。