Clinical Chemistry and Transfusion Medicine, Växjö Central Hospital, Växjö, Sweden.
Department of Research and Development, Region Kronoberg, Växjö, Sweden.
Transfusion. 2019 Oct;59(10):3214-3218. doi: 10.1111/trf.15481. Epub 2019 Aug 9.
Quality control of residual white blood cells (WBCs) and red blood cells (RBCs) in leukoreduced plasma is mandatory. Although technological advances have been made, analysis of quality controls using routine hematology analyzers has not generally been introduced. The aim of this study was to evaluate if the routine hematology analyzer Sysmex XN-10, (Sysmex Nordic ApS) could be used for quality control of residual WBCs and RBCs in leukoreduced plasma.
Linearity, accuracy, and precision were established for two Sysmex XN-10 analyzers using spiked donor plasma. ADAM rWBC (NanoEnTek) and manual counting in the Bürker chamber (NanoEnTek) were reference methods for WBCs and RBCs, respectively. Twenty-five consecutive leukoreduced donor plasma samples were also tested.
For WBCs, the linearity criteria were met for the ADAM rWBC, but not for the Sysmex XN-10 instruments. Precision on both Sysmex XN-10 instruments was accurate only at 6 cells/μL, and accuracy was consistently acceptable only at 5 to 6 cells/μL. The precision and accuracy of the ADAM rWBC were acceptable at 2 to 6 cells/μL. For RBCs, both Sysmex XN-10 instruments and manual counting in the Bürker chamber were linear and fulfilled the precision criteria. Accuracy was acceptable for both Sysmex instruments at 6 to 12 × 10 WBCs/L but fluctuated within the study's measuring range for the Bürker chamber. No false-positive results were seen in the 25 consecutive donor plasma samples tested.
For quality control purposes of leukoreduced plasma, the Sysmex XN-10 analyzer is suitable for the enumeration of residual RBCs but not of residual WBCs.
减少白细胞(WBC)和红细胞(RBC)后的血浆中的白细胞和红细胞的质量控制是强制性的。尽管技术已经取得了进步,但使用常规血液分析仪分析质量控制尚未得到普遍应用。本研究旨在评估 Sysmex XN-10 常规血液分析仪(Sysmex Nordic ApS)是否可用于减少白细胞后的血浆中残留白细胞和红细胞的质量控制。
使用添加供体血浆的方法对两台 Sysmex XN-10 分析仪进行线性度、准确性和精密度评估。ADAM rWBC(NanoEnTek)和 Bürker 室的手动计数(NanoEnTek)分别是 WBC 和 RBC 的参考方法。还测试了 25 个连续的减少白细胞的供体血浆样本。
对于 WBC,ADAM rWBC 的线性度标准得到满足,但 Sysmex XN-10 仪器未得到满足。两台 Sysmex XN-10 仪器的精密度仅在 6 个细胞/μL 时准确,准确性仅在 5 到 6 个细胞/μL 时一致可接受。ADAM rWBC 的精密度和准确性在 2 到 6 个细胞/μL 时可接受。对于 RBC,两台 Sysmex XN-10 仪器和 Bürker 室的手动计数均呈线性且符合精密度标准。两台 Sysmex 仪器的准确性在 6 到 12×10 WBC/L 时可接受,但在 Bürker 室的研究测量范围内波动。在测试的 25 个连续供体血浆样本中未出现假阳性结果。
对于减少白细胞后的血浆的质量控制目的,Sysmex XN-10 分析仪适用于残留 RBC 的计数,但不适用于残留 WBC 的计数。