采用三维动态对比增强超声与二维 DCE-US 评估兔原位肝肿瘤血管生成。

Assessment of angiogenesis in rabbit orthotropic liver tumors using three-dimensional dynamic contrast-enhanced ultrasound compared with two-dimensional DCE-US.

机构信息

Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, The First Affiliated Hospital of Sun Yat-Sen University, No. 58 Zhongshan Road 2, Guangzhou, 510080, People's Republic of China.

Zhongshan School of Medicine, Sun Yat-Sen University, No. 74 Zhongshan Road 2, Guangzhou, 510080, People's Republic of China.

出版信息

Jpn J Radiol. 2019 Oct;37(10):701-709. doi: 10.1007/s11604-019-00861-z. Epub 2019 Aug 10.

Abstract

OBJECTIVES

To evaluate quantitative three-dimensional (3D) dynamic contrast-enhanced ultrasound (DCE-US) in the assessment of tumor angiogenesis using an orthotropic liver tumor model.

METHODS

Nine New Zealand white rabbits with liver orthotropic VX2 tumors were established and imaged by two-dimensional (2D) and 3D DCE-US after SonoVue bolus injections. The intraclass correlation coefficients of perfusion parameters, including peak intensity (PI), mean transit time, time to peak, and area under the curve, were calculated based on time-intensity curve. The percentage area of microvascular (PAMV) and the expression of vascular endothelial growth factor (VEGF) were both evaluated by immunohistochemical analysis and weighted by the tumor activity area ratio. Correlations between quantitative and histologic parameters were analyzed.

RESULTS

The reproducibility of 3D DCE-US quantitative parameters was excellent (ICC 0.91-0.99); but only PI showed high reproducibility (ICC 0.97) in 2D. None of the parameters of quantitative 2D DCE-US were significantly correlated with weighted PAMV or VEGF. For 3D DCE-US, there was a positive correlation between PI and weighted PAMV (r = 0.74, P = 0.04) as well as VEGF (r = 0.79, P = 0.02).

CONCLUSION

Quantitative parameters of 3D DCE-US show feasibility, higher reproducibility and accuracy for the assessment of tumor angiogenesis using an orthotropic liver tumor model compared with 2D DCE-US.

摘要

目的

使用肝原位移植瘤模型评估定量三维(3D)动态对比增强超声(DCE-US)在肿瘤血管生成评估中的作用。

方法

建立 9 只新西兰大白兔肝原位 VX2 肿瘤模型,在 SonoVue 团注后进行二维(2D)和 3D DCE-US 成像。基于时间-强度曲线,计算灌注参数的组内相关系数,包括峰值强度(PI)、平均渡越时间、达峰时间和曲线下面积。微血管面积百分比(PAMV)和血管内皮生长因子(VEGF)的表达均通过免疫组化分析进行评估,并通过肿瘤活性面积比进行加权。分析定量和组织学参数之间的相关性。

结果

3D DCE-US 定量参数的可重复性非常好(ICC 0.91-0.99);但 2D 中仅 PI 具有较高的可重复性(ICC 0.97)。定量 2D DCE-US 的参数与加权 PAMV 或 VEGF 均无显著相关性。对于 3D DCE-US,PI 与加权 PAMV(r=0.74,P=0.04)和 VEGF(r=0.79,P=0.02)呈正相关。

结论

与 2D DCE-US 相比,肝原位移植瘤模型中 3D DCE-US 的定量参数在肿瘤血管生成评估中具有可行性、更高的可重复性和准确性。