Department of Neuro-Ophthalmology (JAM, FDB), National Hospital for Neurology and Neurosurgery, London, United Kingdom; Department of Ophthalmology (JAM), King's College Hospital London, United Kingdom; and Department of Neurosurgery (LD, AKT), National Hospital for Neurology and Neurosurgery, London, United Kingdom.
J Neuroophthalmol. 2020 Jun;40(2):174-177. doi: 10.1097/WNO.0000000000000815.
Assessment of spontaneous venous pulsation (SVP) is commonly undertaken to help determine whether intracranial pressure (ICP) is elevated. Previous studies using direct ophthalmoscopy or slit-lamp assessments have found that SVP is not observed in 67%-81% of subjects with normal ICP, and that interobserver agreement when grading SVP is poor.
Patients (n = 105) undergoing clinically indicated retinal OCT scans, who were all believed to have normal ICP, had 10-second infrared video recordings performed with the Heidelberg Spectralis OCT system (Heidelberg Engineering GmbH, Heidelberg, Germany). The presence and amplitude of SVP in each video was independently graded by 2 neuro-ophthalmologists.
The 2 observers found SVP present in 97% and 98% of right eyes and in one or both eyes in 99% and 100% of subjects. Interobserver agreement was high (Cohen's kappa 0.82 for right eyes). Optic discs with a smaller cup had a significantly lower SVP amplitude (Spearman's rho = 0.22, P = 0.02).
Infrared video is widely available in eye clinics by the use of OCT imaging systems and is substantially more sensitive in detecting SVP than traditional assessments using ophthalmoscopy. SVP is absent in as few as 1% of people with presumed normal ICP.
评估自发性静脉搏动(SVP)通常用于帮助确定颅内压(ICP)是否升高。使用直接检眼镜或裂隙灯评估的先前研究发现,在 67%-81%的正常 ICP 受试者中未观察到 SVP,并且在对 SVP 进行分级时观察者间的一致性较差。
接受临床指示的视网膜 OCT 扫描的患者(n=105),均被认为 ICP 正常,使用海德堡 Spectralis OCT 系统(德国海德堡工程公司,海德堡)进行 10 秒红外视频记录。两名神经眼科医生独立评估每个视频中 SVP 的存在和幅度。
两名观察者发现 97%和 98%的右眼以及 99%和 100%的受试者的一只或两只眼睛存在 SVP。观察者间的一致性很高(右眼的 Cohen's kappa 为 0.82)。杯盘比小的视盘 SVP 幅度明显较低(Spearman's rho=0.22,P=0.02)。
通过 OCT 成像系统,在眼科诊所广泛使用红外视频,在检测 SVP 方面比使用检眼镜的传统评估方法灵敏得多。在被认为 ICP 正常的人中,SVP 缺失的比例低至 1%。