Sapundzhiev Nikolay R, Davidov Georgi, Platikanov Viliyan, Stoyanov George S, Ignatov Valentin
Otolaryngology, Medical University of Varna, Varna, BGR.
Otolaryngology, St. Marina University Hospital, Varna, BGR.
Cureus. 2019 Jul 29;11(7):e5266. doi: 10.7759/cureus.5266.
Introduction The aim of this study was to evaluate the fire risks associated with a harmonic scalpel, with an established avian model simulating oropharyngeal/laryngeal surgery. Methods A standard polyvinyl-chloride (PVC) endotracheal tube (ETT) was inserted into a degutted, whole raw chicken through which 100% oxygen was piped at 10 L/min. The inflated cuff of the tube was grasped and sectioned with the jaws of a standard high-power ultrasonic dissection system Ultracision (Ethicon Endosurgery, Cincinnati, Ohio, USA). Then, the whole ETT was grasped and cut, leaving the device in contact with the ETT for two more minutes. In a second step under the same conditions, an electrosurgical device was placed into the chicken cavity and activated at the chicken tissue near the ETT at a setting of 20 W. All trials were repeated to ensure accuracy. Results No ignition could be produced with the harmonic scalpel under any operation mode settings. In all cases, the ETT was cut through with some fumes and brown discoloration at the site of contact. The electrosurgical device easily caused flash ignition within seconds. Conclusion The harmonic scalpel appears to be a safer tool than electrosurgical devices in the setting of open cavity surgery in oxygen-enriched environments with respect to the presence of flammable medical PVC devices as ETT or catheters.
引言 本研究的目的是使用一个既定的模拟口咽/喉手术的禽类模型,评估与谐波手术刀相关的火灾风险。方法 将一根标准的聚氯乙烯(PVC)气管内插管(ETT)插入一只去内脏的整只生鸡体内,以10升/分钟的速度向其输送100%的氧气。用标准的高功率超声解剖系统Ultracision(美国俄亥俄州辛辛那提市Ethicon Endosurgery公司生产)的钳口抓住并切开气管内插管充气的套囊。然后,抓住并切断整个气管内插管,使该设备与气管内插管接触两分钟。在第二步中,在相同条件下,将一个电外科设备放入鸡腔内,并在气管内插管附近的鸡组织处以20瓦的功率启动。所有试验均重复进行以确保准确性。结果 在任何操作模式设置下,谐波手术刀都不会引发着火。在所有情况下,气管内插管都被切断,接触部位有一些烟雾和褐色变色。电外科设备在几秒钟内很容易引发闪燃。结论 就存在作为气管内插管或导管的易燃医用PVC设备而言,在富氧环境下的开放腔手术中,谐波手术刀似乎比电外科设备更安全。