经颅磁刺激安全性:从操作人员暴露角度来看。

Transcranial magnetic stimulation safety from operator exposure perspective.

机构信息

Graduate Program in Biomedical Engineering, University of Manitoba, Winnipeg, MB, R3T 5V6, Canada.

Department of Electrical & Computer Engineering, University of Manitoba, Winnipeg, MB, R3T 5V6, Canada.

出版信息

Med Biol Eng Comput. 2020 Feb;58(2):249-256. doi: 10.1007/s11517-019-02084-w. Epub 2019 Dec 13.

Abstract

A simulated model of a commercial transcranial magnetic stimulation (TMS) coil is analyzed to determine electromagnetic field (EMF) exposure for an operator while holding or adjusting the coil. Induced EMF strengths are calculated using a commercial figure-8 coil geometry and pulse configuration, with geometrical representations of the subject's head and the operator's head, torso, and hand. Exposure levels are compared to experimental results in the literature and international guidelines for occupational EMF exposure limits. Exposure limit guidelines of 0.8 V/m rms are exceeded at approximately 24.6 cm from the coil for the torso model and at 20.3 cm for the head model measured perpendicular to the plane of the coil. In the plane of the coil, the operator can approach closer without exceeding guidelines. The results in the hand model along the edge of the coil give 9.9 V/m and 88.5 V/m for average and peak field strength, respectively. A discussion of the potential consequences of operator exposure to fields exceeding published guidelines concludes that since the guidelines are only concerned with acute effects and do not suggest any potential chronic effects, occupational exposure in the context of delivering TMS treatment may be considered reasonable. Graphical abstract A model of an operator's head/torso was moved in space relative to a standard TMS coil and subject. Positions at which safety guidelines are exceeded were calculated. The maximum induced electric field was also calculated in a hand model placed in a position commonly used to hold TMS coils during treatments.

摘要

对商业经颅磁刺激(TMS)线圈的模拟模型进行分析,以确定操作人员握持或调整线圈时的电磁场(EMF)暴露情况。使用商业 figure-8 线圈几何形状和脉冲配置计算感应 EMF 强度,并对受试者头部和操作人员头部、躯干和手部的几何形状进行表示。将暴露水平与文献中的实验结果和职业 EMF 暴露限值的国际指南进行比较。对于躯干模型,在距线圈约 24.6 厘米处和对于头部模型,在垂直于线圈平面处的 20.3 厘米处,超过了 0.8 V/m rms 的暴露限值指南。在线圈平面内,操作人员可以更接近而不会超过指南。在线圈边缘的手部模型中,平均和峰值场强分别为 9.9 V/m 和 88.5 V/m。讨论操作人员暴露于超过公布指南的场的潜在后果得出的结论是,由于指南仅关注急性影响,并且不建议任何潜在的慢性影响,因此在提供 TMS 治疗的背景下,职业暴露可能被认为是合理的。 图形摘要 操作人员的头部/躯干模型相对于标准 TMS 线圈和受试者在空间中移动。计算了超过安全指南的位置。还在治疗过程中通常用于握持 TMS 线圈的位置处的手部模型中计算了最大感应电场。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验