Suppr超能文献

使用新型便携式设备对肌肉损伤患者进行等长膝关节伸展力量的常规监测:与标准等速测力计的交叉验证。

Routine monitoring of isometric knee extension strength in patients with muscle impairments using a new portable device: cross-validation against a standard isokinetic dynamometer.

机构信息

Institute of Myology, Neuromuscular Investigation Center, Pitié-Salpêtrière University Hospital, Paris, France. Author to whom any correspondence should be addressed.

出版信息

Physiol Meas. 2020 Feb 5;41(1):015003. doi: 10.1088/1361-6579/ab6b49.

Abstract

OBJECTIVE

Muscle strength is a critical clinical hallmark in both health and disease. The current study introduces a novel portable device prototype (MyoQuad) for assessing and monitoring maximal voluntary isometric knee extension torque (MVIT).

APPROACH

Fifty-six patients with inclusion body myositis were studied. Knee extension weakness is a key feature in this inflammatory muscle disease. Cross-validation with an isokinetic dynamometer (Biodex System 3 Pro) was performed. Between-day reproducibility and ability to monitor changes in muscle strength over time compared to the gold standard method as a reference, were also investigated.

MAIN RESULTS

The measurement was feasible even in the weakest patients. Agreement between methods was excellent (standard error of measurement (SEM) was 3.8 Nm and intra-class correlation coefficient (ICC) was 0.973). Least significant difference (LSD) was 4.9 and 5.3 Nm for the MyoQuad and the Biodex, respectively Measurements using the MyoQuad exhibited excellent between-day reproducibility (SEM was 2.4 Nm and ICC was 0.989 versus 2.6 Nm and 0.988 using the Biodex). Changes in MVIT at 6 and 12 months were similar between methods (timepoint  ×  method interaction was not significant; all p   >  0.19); strength changes classified according to LSD at 6 and 12 months were consistent between methods (>70% consistent classification)).

SIGNIFICANCE

The measurement of MVIT using the MyoQuad offers a cost-effective, portable and immediate alternative for the routine measurement of maximal voluntary isometric strength of the quadriceps. The MyoQuad offers a comfort and stability that cannot be provided by standard hand-held dynamometers. These results support quantitative muscle strength assessment using fixed yet flexible dynamometry within clinical routine and multicenter trials.

摘要

目的

肌肉力量是健康和疾病的重要临床特征。本研究介绍了一种新的便携式设备原型(MyoQuad),用于评估和监测最大自主等长膝关节伸展扭矩(MVIT)。

方法

对 56 名包涵体肌炎患者进行了研究。膝关节伸展无力是这种炎症性肌肉疾病的一个关键特征。与等速测力计(Biodex System 3 Pro)进行了交叉验证。还研究了与金标准方法作为参考相比,该方法的日内重复性和监测肌肉力量随时间变化的能力。

主要结果

即使在最弱的患者中,该测量也是可行的。两种方法之间的一致性非常好(测量误差为 3.8 Nm,组内相关系数为 0.973)。MyoQuad 和 Biodex 的最小可检测差异(LSD)分别为 4.9 和 5.3 Nm。MyoQuad 的测量具有极好的日内可重复性(测量误差为 2.4 Nm,组内相关系数为 0.989,而 Biodex 分别为 2.6 Nm 和 0.988)。两种方法在 6 个月和 12 个月时的 MVIT 变化相似(时间点与方法的交互作用不显著;所有 p 值均>0.19);根据 6 个月和 12 个月时的 LSD 分类的力量变化在两种方法之间是一致的(>70%的一致性分类))。

意义

使用 MyoQuad 测量 MVIT 为常规测量股四头肌最大自主等长力量提供了一种经济实惠、便携和即时的替代方法。MyoQuad 提供了标准手持测力计无法提供的舒适性和稳定性。这些结果支持在临床常规和多中心试验中使用固定但灵活的测力计进行定量肌肉力量评估。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验