Piatin V F
Biull Eksp Biol Med. 1988 Nov;106(11):519-21.
The article describes a simple floating hydraulic micromanipulator of original construction which works in positive-positive pressure and provides maximum stability of the microelectrode tip concerning the neuron in the moving brain. This work is achieved by fixing of the micromanipulator on the hinged suspension bracket and by setting of it on the surface of the object with the help of bearing site. Breathing and vascular brain moving don't influence the neuron activity lead.
本文描述了一种结构新颖的简单浮动式液压显微操作器,它在正-正压力下工作,能使微电极尖端相对于运动中的大脑中的神经元保持最大稳定性。这一成果是通过将显微操作器固定在铰接式悬挂支架上,并借助承载部位将其放置在物体表面来实现的。呼吸和大脑血管的运动不会影响神经元活动导线。