Gough N A, Morgan G, McKenna D M, Hibbard B M, Newcombe R G
Department of Obstetrics, University of Wales College of Medicine, UK.
Clin Phys Physiol Meas. 1988 Nov;9(4):371-4. doi: 10.1088/0143-0815/9/4/009.
Details of the design of obstetric forceps are important to ensure that they can be used safely for mother and baby. One of the most important measurements is the radius of the cephalic curve of the blades since if this is inappropriate there will be a poor grip on the baby's head and risk of slippage, superficial injury or even intracranial haemorrhage and brain damage because of undue compression. Because of the complex design of the forceps blades, which have a pelvic curve as well as a cephalic curve and a fenestra, the radius of the cephalic curve is difficult to measure. A device was designed to facilitate this measurement and was found to be accurate and reliable in use. This was the first stage in the study to assess the suitability for modern practice of instruments designed many years ago.
产科产钳的设计细节对于确保其能安全用于母婴至关重要。最重要的测量指标之一是钳叶头部曲线的半径,因为如果该半径不合适,就会对胎儿头部抓握不佳,存在滑脱风险、造成体表损伤,甚至因过度压迫导致颅内出血和脑损伤。由于钳叶设计复杂,既有骨盆曲线、头部曲线,还有开窗,所以头部曲线的半径很难测量。为此设计了一种装置来便于进行此项测量,且经实际使用发现该装置准确可靠。这是本研究评估多年前设计的器械是否适用于现代临床实践的第一阶段。