单日应激-静息心肌灌注成像方案中最佳 Tc 放射性活度比:一项多探头 SPECT 体模研究。

Optimal Tc activity ratio in the single-day stress-rest myocardial perfusion imaging protocol: A multi-SPECT phantom study.

机构信息

Department of Nuclear Medicine, S. Maugeri Foundation, IRCCS, Scientific Institute of Veruno (NO), Veruno, Italy.

Department of Medical Physics, University Hospital 'Maggiore della Carità', Novara, Italy.

出版信息

J Nucl Cardiol. 2021 Feb;28(1):338-349. doi: 10.1007/s12350-020-02290-2. Epub 2020 Jul 27.

Abstract

BACKGROUND

This investigation used image data generated by an anthropomorphic phantom to determine the minimal Tc rest-stress activity concentration ratio (R) able to minimize the ghosting effect in the single-day stress-first myocardial perfusion imaging, using different positions of the perfusion defect (PD), scanners and reconstruction protocols.

METHODS

A cardiac phantom with a simulated PD was imaged under different R using different gamma cameras and reconstruction algorithms. The residual activity from precedent stress administration was simulated by modeling effective half-times in each compartment of the phantom and assuming a delay of 3 hours between the stress and rest studies. The net contrast (NC) of the PD in the rest study was assessed for different R, PD positions and scanner/software combinations. The optimal R will be the one that minimize the NC in the rest images RESULTS: The activity concentration ratio R, the position of the PD and the scanner/software combinations were all main effects with a statistically significant impact on the NC, in decreasing order of relevance. The NC diminished significantly only for R values up to 2. No further improvement was observed for NC for R values above 2 and up to 3. NC was significantly higher in anteroseptal than in posterolateral positions of the PD and higher for solid-state cameras.

CONCLUSIONS

A rest-stress activity concentration ratio R of 2 in single-day stress-first myocardial perfusion imaging is enough to achieve the maximum net contrast in the PD. This ratio should be used to optimize patient's radiation exposure.

摘要

背景

本研究利用人体模型体模生成的图像数据,确定了最小 Tc 静息-应激活性浓度比(R),以最小化单日内首次负荷心肌灌注成像中的重影效应,同时考虑了不同位置的灌注缺损(PD)、扫描仪和重建方案。

方法

采用不同的伽马相机和重建算法,对具有模拟 PD 的心脏体模在不同 R 下进行成像。通过对体模各腔室的有效半衰期进行建模,并假设在应激和静息研究之间存在 3 小时的延迟,模拟了先前应激给药的残留活性。评估了不同 R、PD 位置和扫描仪/软件组合下 PD 在静息研究中的净对比(NC)。最佳 R 将是使静息图像中 NC 最小化的 R。

结果

R、PD 位置和扫描仪/软件组合都是对 NC 有统计学显著影响的主要因素,其相关性依次降低。只有 R 值高达 2 时,NC 才会显著降低。对于 R 值大于 2 至 3 的 NC,未观察到进一步的改善。PD 在前间隔位置的 NC 明显高于后外侧位置,固态相机的 NC 更高。

结论

在单日内首次负荷心肌灌注成像中,R 为 2 的静息-应激活性浓度比足以使 PD 获得最大净对比。该比值可用于优化患者的辐射暴露。