[在CT透视中使用半扫描降低操作员暴露辐射剂量及图像质量评估]

[Reduction of Operator Exposure Radiation Dose and Evaluation of Image Quality Using Half Scan in CT Fluoroscopy].

作者信息

Kinoshita Emi, Miyazaki Hitoshi, Ogawa Kazuhisa, Komiya Isao, Kato Toyoyuki

机构信息

Department of Radiology, Division of Medical Technology, Kyushu University Hospital.

出版信息

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2020;76(12):1266-1275. doi: 10.6009/jjrt.2020_JSRT_76.12.1266.

Abstract

In recent years, the number of examinations and treatments using computed tomography fluoroscopy (CTF) has been increasing, and there is concern about an increase in the exposure radiation dose of the operator. Use of half scan CTF can be expected to reduce the exposure radiation dose, but there is no report. The purpose of this study was to evaluate the exposure radiation dose at the operator's position and image quality when using a half scan CTF. The left side facing the gantry was the operator's position, and the ambient dose equivalent at 160 cm, 130 cm, and 100 cm from the floor was measured using an ionization chamber survey meter. The absorbed dose at the forceps holding position of the operator was measured using a fluorescent glass dosimeter with the forceps holding position 15 cm caudal from the scan center. The imaging conditions used a tube voltage of 120 kV and a tube current of 50 mA. Half scan CTF was performed by changing the center angle of the half scan on the console every 45°. As a result, the set angles were 135°and 90°at the operator's position, and 135°at the operator's forceps holding position. In addition, we evaluated the effect of half scan CTF on image quality. CTF images were collected with a cryogenic needle used for cryotherapy punctured in a water-equivalent self-made phantom. The profile curves of the obtained images were drawn and compared using analysis software to evaluate the effects of artifacts. Then, the SD of the CT value of the region of interest with and without the artifact was measured, and the relative artifact index was calculated and evaluated. Using the same image, CT value and SD were tested to evaluate noise. Half scan CTF had no effect on the image quality due to artifacts and noise.

摘要

近年来,使用计算机断层扫描透视(CTF)进行的检查和治疗数量不断增加,人们担心操作人员的照射辐射剂量会增加。使用半扫描CTF有望降低照射辐射剂量,但尚无相关报告。本研究的目的是评估使用半扫描CTF时操作人员位置的照射辐射剂量和图像质量。面对机架的左侧为操作人员位置,使用电离室巡测仪测量距地面160 cm、130 cm和100 cm处的周围剂量当量。使用荧光玻璃剂量计测量操作人员持钳位置的吸收剂量,持钳位置位于扫描中心尾侧15 cm处。成像条件为管电压120 kV,管电流50 mA。通过在控制台上每隔45°改变半扫描的中心角度来进行半扫描CTF。结果,操作人员位置的设定角度为135°和90°,操作人员持钳位置的设定角度为135°。此外,我们评估了半扫描CTF对图像质量的影响。使用用于冷冻治疗穿刺的低温针在自制的水等效体模中穿刺来采集CTF图像。使用分析软件绘制并比较获得图像的轮廓曲线,以评估伪影的影响。然后,测量有无伪影的感兴趣区域的CT值标准差,计算并评估相对伪影指数。使用相同图像,测试CT值和标准差以评估噪声。半扫描CTF对因伪影和噪声导致的图像质量没有影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验