Impact of Trifocal and Trifocal Toric Intraocular Lenses on Spectral-domain OCT Retinal Measurements.

机构信息

Clínica Rementería.

Ramón y Cajal University Hospital, Ramón y Cajal Institute for Health Research (IRYCIS).

出版信息

J Glaucoma. 2021 Apr 1;30(4):300-303. doi: 10.1097/IJG.0000000000001743.

Abstract

PRECIS

Monofocal, trifocal, and trifocal toric intraocular lenses (IOLs) induce similar changes on overall retinal optical coherence tomography (OCT) measurements.

PURPOSE

The purpose of this study was to assess whether trifocal and trifocal toric IOLs affect the measurement of retinal parameters using spectral-domain OCT.

METHODS

This cross-over study included patients undergoing implantation of a trifocal IOL: AcrySof IQ PanOptix, a trifocal toric IOL: AcrySof IQ PanOptix Toric, and a monofocal IOL: AcrySof IQ. The monofocal group was considered as the control group. The refractive target was emmetropia in all cases. Mean average macular thickness, macular volume, and retinal nerve fiber layer (RNFL) thickness were measured with the Cirrus HD-OCT. All measurements were performed before and 3 months after the surgery.

RESULTS

The study analyzed 150 eyes of 150 patients (50 for each IOL group). Macular thickness and macular volume showed statistically significant differences before and after the surgery for the 3 groups (P<0.05 in all cases). RNFL thickness was found to be similar before and after the surgery in all groups (P>0.05 in all cases). Mean difference values (before and after the surgery) in the monocular, trifocal, and trifocal toric group for macular thickness, macular volume, and RNFL thickness were 4.9±7.8, 7.9±10.0, and 7.7±13.7 µm, respectively; 0.1±0.2, 0.2±0.4, and 0.2±0.3 mm3, respectively; and 0.8±5.5, 1.3±6.0, and 0.8±6.7 µm, respectively. Mean differences were found to be similar for the 3 groups.

CONCLUSION

The trifocal and the trifocal toric IOLs under study did not induce an additional impact on spectral-domain OCT retinal measurements compared with monofocal IOLs.

摘要

摘要

单焦点、三焦点和三焦点散光人工晶状体(IOL)在整体视网膜光相干断层扫描(OCT)测量中引起相似的变化。

目的

本研究旨在评估三焦点和三焦点散光 IOL 是否会影响光谱域 OCT 对视网膜参数的测量。

方法

这是一项交叉研究,包括植入三焦点 IOL(AcriSof IQ PanOptix)、三焦点散光 IOL(AcriSof IQ PanOptix Toric)和单焦点 IOL(AcriSof IQ)的患者。单焦点组被视为对照组。所有情况下的屈光目标均为正视。使用 Cirrus HD-OCT 测量平均黄斑厚度、黄斑体积和视网膜神经纤维层(RNFL)厚度。所有测量均在手术前和手术后 3 个月进行。

结果

该研究分析了 150 名患者的 150 只眼(每组 50 只眼)。三组患者的黄斑厚度和黄斑体积在手术前后均有统计学差异(所有情况下 P<0.05)。所有组在手术前后的 RNFL 厚度均相似(所有情况下 P>0.05)。单眼、三焦点和三焦点散光组黄斑厚度、黄斑体积和 RNFL 厚度的平均差值(手术前后)分别为 4.9±7.8、7.9±10.0 和 7.7±13.7μm;0.1±0.2、0.2±0.4 和 0.2±0.3mm3;0.8±5.5、1.3±6.0 和 0.8±6.7μm。三组之间的平均差值相似。

结论

与单焦点 IOL 相比,研究中的三焦点和三焦点散光 IOL 对光谱域 OCT 视网膜测量没有产生额外的影响。