Aging Clin Exp Res. 2021 Nov;33(11):2989-2994. doi: 10.1007/s40520-021-01832-z. Epub 2021 Mar 28.
Gait speed estimation using wearable inertial sensors during daily activities suffers from high complexity and inaccuracies in distance estimation when integrating acceleration signals. The aim of the study was to investigate the agreement between the methods of gait speed estimation using the persons' walk ratio (step-length/step-frequency relation) or step-frequency (number of steps per minute) and a "gold standard".
For this cross-sectional validation study, 20 healthy community-dwelling older persons (mean age 72.1 years; 70% women) walked at slow, normal, and fast speed over an instrumented walkway (reference measure). Gait speed was calculated using the person's pre-assessed walk ratio. Furthermore, the duration of walking and number of steps were used for calculation.
The agreement between gait speed calculation using the walk ratio or step-frequency (adjusted to body height) and reference was r = 0.98 and r = 0.93, respectively. Absolute and relative mean errors of calculated gait speed using pre-assessed walk ratio ranged between 0.03-0.07 m/s and 1.97-4.17%, respectively.
After confirmation in larger cohorts of healthy community-dwelling older adults, the mean gait speed of single walking bouts during activity monitoring can be estimated using the person's pre-assessed walk ratio. Furthermore, the mean gait speed can be calculated using the step-frequency and body height and can be an additional parameter in stand-alone activity monitoring.
在日常活动中使用可穿戴惯性传感器估计步态速度时,在整合加速度信号时,距离估计的复杂性和准确性都很高。本研究的目的是研究使用个体的步长/步频比(步长与步频的关系)或步频(每分钟的步数)来估计步态速度的方法与“金标准”之间的一致性。
本横断面验证研究纳入了 20 名健康的社区居住的老年人(平均年龄 72.1 岁;70%为女性),他们在仪器化步道上以慢、中、快三种速度行走(参考测量)。使用个体预先评估的步长比来计算步态速度。此外,还使用行走的时长和步数来进行计算。
使用步长比或步频(校正至身高)计算的步态速度与参考值之间的一致性分别为 r=0.98 和 r=0.93。使用预先评估的步长比计算的步态速度的绝对和相对平均误差分别在 0.03-0.07 m/s 和 1.97-4.17%之间。
在更大的健康社区居住的老年人队列中得到验证后,活动监测中单个行走回合的平均步态速度可以使用个体预先评估的步长比来估计。此外,可以使用步频和身高来计算平均步态速度,并且可以成为独立活动监测的附加参数。