组件尺寸和校准误差对全髋关节置换数字化模板的影响。来自简单数学模型的直观矩阵。

Implications of component size and calibration error on digital templating for total hip arthroplasty. A visual matrix from a simple mathematical model.

机构信息

Department of Orthopaedic and Trauma Surgery, University Hospital of Cologne, Cologne, Germany.

Division of Orthopaedics, Department of Trauma and Orthopaedic Surgery, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.

出版信息

Int J Comput Assist Radiol Surg. 2021 Jun;16(6):1037-1041. doi: 10.1007/s11548-021-02367-4. Epub 2021 Apr 17.

Abstract

OBJECTIVE

Preoperative digital templating is a standard procedure in total hip arthroplasty. Deviations between template size and final implant size may result from inaccurate calibration, templating as well as intraoperative decisions. So far, the explicit effect of calibration errors on templating has not been addressed adequately.

MATERIALS AND METHODS

A mathematical simulation of calibration errors up to ± 24% was applied to the templating of acetabular cups (38 to 72 mm diameter). The effect of calibration errors on template component size as deviation from optimal size was calculated.

RESULTS

The relationship between calibration error and component size deviation is inverse and linear. Calibration errors have a more pronounced effect on larger component sizes. Calibration errors of 2-6% result in templating errors of up to two component sizes. Common errors of up to 12% may result in templating errors of 3-4 sizes for common implant sizes. A tabular matrix visualizes the effect.

CONCLUSION

Calibration errors play a significant role in component size selection during digital templating. Orthopedic surgeons should be aware of this effect and try to identify and address this source of error.

摘要

目的

术前数字化模板是全髋关节置换术的标准程序。由于校准、模板制作以及术中决策的不精确,模板尺寸与最终植入物尺寸之间可能存在偏差。迄今为止,校准误差对模板制作的明确影响尚未得到充分解决。

材料和方法

对髋臼杯(直径 38 至 72 毫米)的模板进行高达±24%的校准误差的数学模拟。计算校准误差对模板组件尺寸与最佳尺寸偏差的影响。

结果

校准误差与组件尺寸偏差之间呈反比线性关系。校准误差对较大组件尺寸的影响更为显著。2-6%的校准误差会导致模板误差高达两个组件尺寸。常见的误差高达 12%可能会导致常见植入物尺寸的模板误差为 3-4 个尺寸。一个表格矩阵可视化了这种影响。

结论

校准误差在数字化模板制作过程中对组件尺寸选择起着重要作用。矫形外科医生应该意识到这一影响,并尝试识别和解决这一误差源。