Department of Orthopaedic Medical Engineering, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita, Osaka, 565-0871, Japan.
Department of Orthopaedic Surgery, Osaka University Graduate School of Medicine, Suita, Osaka, Japan.
Int Orthop. 2021 Jul;45(7):1719-1725. doi: 10.1007/s00264-021-05015-3. Epub 2021 Apr 20.
We investigated the accuracy of the acetabular cup position and orientation in robotic-assisted total hip arthroplasty (rTHA) compared to navigated THA (nTHA) using computed tomography (CT) for patients with osteoarthritis secondary to developmental dysplasia of the hip (DDH).
We studied 31 hips of 28 patients who underwent rTHA and 119 hips of 112 patients who underwent nTHA with the same target cup orientation. After propensity score matching, each group comprised 29 hips. Post-operative cup position and orientation were measured from the postoperative CT data. Errors from the target cup position and orientation were compared between the two groups.
Post-operatively, the absolute error of the anteroposterior and superoinferior cup positions from the target position was significantly lower in the rTHA group than in the nTHA group. The change in the post-operative radiographic inclination from the target orientation was lower in the rTHA group than in the nTHA group. Screw fixation for cup implantation was required for three hips in the nTHA group but not in the rTHA group.
rTHA achieved more precise cup implantation with reduced variation from the target orientation compared to nTHA in patients with osteoarthritis secondary to DDH.
我们通过计算机断层扫描(CT)比较了机器人辅助全髋关节置换术(rTHA)与导航辅助全髋关节置换术(nTHA)在髋臼杯位置和方向的准确性,研究对象为继发于髋关节发育不良(DDH)的骨关节炎患者。
我们研究了 28 名患者的 31 髋接受 rTHA 治疗,112 名患者的 119 髋接受 nTHA 治疗,两组患者髋臼杯的目标方向相同。在进行倾向评分匹配后,每组各包含 29 髋。从术后 CT 数据测量术后髋臼杯的位置和方向。比较两组之间目标髋臼杯位置和方向的误差。
术后,rTHA 组髋臼杯前后位和上下位的绝对误差明显低于 nTHA 组。rTHA 组术后影像学倾斜角度的变化小于 nTHA 组。nTHA 组中有 3 髋需要螺钉固定髋臼杯,而 rTHA 组则不需要。
与 nTHA 相比,rTHA 可实现更精确的髋臼杯植入,髋臼杯方向的变化更小,尤其在继发于 DDH 的骨关节炎患者中。