Polytetrafluoroethylene coating fragments during neuroendovascular therapy: An analysis of two damaged microguidewires.

机构信息

Department of Radiology, University Hospital Rigshospitalet, Copenhagen, Denmark.

Department of Radiology, Næstved-Slagelse Hospitals, Denmark.

出版信息

Interv Neuroradiol. 2022 Feb;28(1):16-21. doi: 10.1177/15910199211015127. Epub 2021 May 26.

Abstract

Cerebral polymer coating embolism from intravascular devices represents a potentially serious complication to endovascular therapy (EVT). We report two cases of neuroendovascular treatment where filamentous polymer fragments were noted possibly due to damage of the surface coating during manipulation and backloading of microguidewires. As the exact origin of the debris was initially not known, microguidewires and fragments were examined with light microscopy, stereomicroscopy, scanning electron microscopy and attenuated-total-reflection Fourier transform infrared spectroscopy. Fragments consisted of polytetrafluoroethylene and silicone oil stemming from the proximal shaft of a standard microguidewire. To our knowledge, this is the first report of polytetrafluoroethylene coating fragments created during EVT. Future studies should assess the mechanism of polymer coating delamination and its potential consequences during EVT including inadvertent fragment migration into the cerebral circulation.

摘要

血管内装置引起的脑聚合物涂层栓塞是血管内治疗(EVT)的一种潜在严重并发症。我们报告了两例神经血管治疗的病例,在微导丝的操作和重新加载过程中,可能由于表面涂层的损坏而出现丝状聚合物碎片。由于最初不知道碎片的确切来源,因此使用光学显微镜、立体显微镜、扫描电子显微镜和衰减全反射傅里叶变换红外光谱法对微导丝和碎片进行了检查。碎片由聚四氟乙烯和硅油组成,源自标准微导丝的近端轴。据我们所知,这是首次在 EVT 过程中报告聚四氟乙烯涂层碎片的情况。未来的研究应评估聚合物涂层分层的机制及其在 EVT 过程中的潜在后果,包括意外的碎片迁移到大脑循环中。

相似文献

1
Polytetrafluoroethylene coating fragments during neuroendovascular therapy: An analysis of two damaged microguidewires.
Interv Neuroradiol. 2022 Feb;28(1):16-21. doi: 10.1177/15910199211015127. Epub 2021 May 26.
2
Fragmentation of Hydrophilic Guidewire Coatings During Neuroendovascular Therapy.
Clin Neuroradiol. 2023 Sep;33(3):793-799. doi: 10.1007/s00062-023-01283-1. Epub 2023 Apr 25.
4
Polymer coating embolism from intravascular medical devices - a clinical literature review.
Cardiovasc Pathol. 2017 Sep-Oct;30:45-54. doi: 10.1016/j.carpath.2017.06.004. Epub 2017 Jun 22.
8
: Intravascular Devices with a Higher Risk of Polymer Emboli: The Need for Particulate Generation Testing.
Biomed Instrum Technol. 2020 Jan-Feb;54(1):37-43. doi: 10.2345/0899-8205-54.1.37.
9
XPS and FTIR Studies of Polytetrafluoroethylene Thin Films Obtained by Physical Methods.
Polymers (Basel). 2019 Oct 9;11(10):1629. doi: 10.3390/polym11101629.
10
Anticoagulation and endothelial cell behaviors of heparin-loaded graphene oxide coating on titanium surface.
Mater Sci Eng C Mater Biol Appl. 2016 Jun;63:333-40. doi: 10.1016/j.msec.2016.03.001. Epub 2016 Mar 3.

本文引用的文献

1
Delayed Complications Due to Polymer Coating Embolism after Endovascular Treatment.
NMC Case Rep J. 2019 Dec 18;7(1):5-10. doi: 10.2176/nmccrj.cr.2018-0319. eCollection 2020 Jan.
4
Polymer coating embolism from intravascular medical devices - a clinical literature review.
Cardiovasc Pathol. 2017 Sep-Oct;30:45-54. doi: 10.1016/j.carpath.2017.06.004. Epub 2017 Jun 22.
6
Foreign Body Emboli following Cerebrovascular Interventions: Clinical, Radiographic, and Histopathologic Features.
AJNR Am J Neuroradiol. 2015 Nov;36(11):2121-6. doi: 10.3174/ajnr.A4415. Epub 2015 Aug 20.
7
Distal embolization of hydrophilic-coating material from coronary guidewires after percutaneous coronary interventions.
Circ Cardiovasc Interv. 2015 Feb;8(2):e001816. doi: 10.1161/CIRCINTERVENTIONS.114.001816.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验