Payne S R, Ramsay J W
Academic Unit, Institute of Urology, London, England.
J Urol. 1988 Sep;140(3):637-41. doi: 10.1016/s0022-5347(17)41745-9.
Intra-pelvic pressure and peristaltic activity were measured before and during diuresis in 26 minipigs, immediately, one week and three weeks following ureteric intubation with 5 and 6F polyethylene double J stents. Changes in these dynamic parameters were seen which indicate that these stents caused renal outflow obstruction. The severity of this obstruction appears dependent upon the size of the catheter and the temporal length of intubation. This study indicates that double J stents provide sub-optimal drainage of the renal pelvis and damaged ureter.
在26只小型猪中,于利尿前及利尿期间,以及在分别用5F和6F聚乙烯双J支架进行输尿管插管后即刻、1周和3周时,测量盆腔内压力和蠕动活性。观察到这些动态参数发生了变化,这表明这些支架导致了肾流出道梗阻。这种梗阻的严重程度似乎取决于导管的大小和插管的时间长度。本研究表明,双J支架对肾盂和受损输尿管的引流效果欠佳。