Abstract

The accuracy of the implant's post-operative position and orientation in reverse shoulder arthroplasty is known to play a significant role in both clinical and functional outcomes. Whilst technologies such as navigation and robotics have demonstrated superior radiological outcomes in many fields of surgery, the impact of augmented reality (AR) assistance in the operating room is still unknown. Malposition of the glenoid component in shoulder arthroplasty is known to result in implant failure and early revision surgery. The use of AR has many promising advantages, including allowing the detailed study of patient-specific anatomy without the need for invasive procedures such as arthroscopy to interrogate the joint's articular surface. In addition, this technology has the potential to assist surgeons intraoperatively in aiding the guidance of surgical tools. It offers the prospect of increased component placement accuracy, reduced surgical procedure time, and improved radiological and functional outcomes, without recourse to the use of large navigation or robotic instruments, with their associated high overhead costs. This feasibility study describes the surgical workflow from a standardised CT protocol, via 3D reconstruction, 3D planning, and use of a commercial AR headset, to AR-assisted k-wire placement. Post-operative outcome was measured using a high-resolution laser scanner on the patient-specific 3D printed bone. In this proof-of-concept study, the discrepancy between the planned and the achieved glenoid entry point and guide-wire orientation was approximately 3 mm with a mean angulation error of 5°.

摘要

已知在反肩关节置换术中,植入物术后的位置和方向准确性在临床和功能结果中都起着重要作用。虽然导航和机器人技术等在许多外科领域已显示出卓越的放射学结果,但手术室中增强现实(AR)辅助的影响仍不明确。肩关节置换术中关节盂部件位置不当会导致植入物失败和早期翻修手术。AR的使用有许多潜在优势,包括无需关节镜等侵入性程序来检查关节表面,就能详细研究患者特定解剖结构。此外,这项技术有潜力在术中协助外科医生引导手术工具。它有望提高部件放置准确性、缩短手术时间,并改善放射学和功能结果,而无需使用大型导航或机器人器械及其相关的高昂费用。这项可行性研究描述了从标准化CT协议,经三维重建、三维规划以及使用商用AR头戴设备,到AR辅助克氏针放置的手术流程。使用针对患者特定的三维打印骨骼的高分辨率激光扫描仪测量术后结果。在这项概念验证研究中,计划的和实际的关节盂入口点及导丝方向之间的差异约为3毫米,平均角度误差为5°。