Abstract

INTRODUCTION

Body fluid cell counting and differentiation provide essential information for diagnosis and monitoring of diverse pathologies. We evaluated the performance of the newly launched Abbott Alinity hq hematology analyzer for automated cell counting in body fluids and compared red blood cell (RBC) and total nucleated cell (TNC) counts with the Cell-Dyn Sapphire automated hematology analyzer. Differential counts were compared with microscopic differentiation on cytocentrifuged preparations.

METHODS

Background concentration limits, limit of detection (LOD), linearity, imprecision, functional sensitivity and carryover were evaluated. For method comparison, we collected 172 body fluids (17 continuous ambulatory peritoneal dialysis fluids, 56 cerebrospinal fluids and 99 serous fluids).

RESULTS

Background concentration limits were ≤1000 cells/μL for RBC counts and ≤3 cells/μL for TNC counts. The LOD was 1000 RBC/µL and 5 TNC/µL. Results from linear regression analysis revealed excellent linearity. Functional sensitivity was 3000 cells/µL for RBC counts and 50 cells/µL for TNC counts. Carryover was 0.6% and 0.1% for TNC and RBC, respectively. The Alinity hq shows good clinical performance.

CONCLUSION

We demonstrated comparable performance for body fluid cell counting between the Alinity hq analyzer and the Cell-Dyn Sapphire. The Alinity hq can be very useful as a screening tool for body fluid cell counting.

摘要

简介

体液细胞计数和分类为各种病理诊断和监测提供了重要信息。我们评估了雅培 Alinity hq 血液分析仪在体液自动细胞计数方面的性能,并与 Cell-Dyn Sapphire 自动化血液分析仪比较了红细胞(RBC)和总核细胞(TNC)计数。将差异计数与细胞离心制备的显微镜差异进行了比较。

方法

评估了背景浓度限制、检测限(LOD)、线性、精密度、功能灵敏度和携带污染。为了进行方法比较,我们收集了 172 份体液(17 份连续流动腹膜透析液、56 份脑脊液和 99 份浆膜液)。

结果

RBC 计数的背景浓度限制为≤1000 个/μL,TNC 计数的背景浓度限制为≤3 个/μL。LOD 为 1000 RBC/µL 和 5 TNC/µL。线性回归分析结果显示具有极好的线性。RBC 计数的功能灵敏度为 3000 个/μL,TNC 计数的功能灵敏度为 50 个/μL。携带污染分别为 TNC 和 RBC 的 0.6%和 0.1%。Alinity hq 具有良好的临床性能。

结论

我们证明了 Alinity hq 分析仪和 Cell-Dyn Sapphire 之间在体液细胞计数方面具有相当的性能。Alinity hq 可以作为体液细胞计数的筛选工具非常有用。