Tumor Immunology Lab, Radboud Institute for Molecular Life Sciences, Radboudumc, Nijmegen, the Netherlands.
Department of Medical Imaging, Radboudumc, Nijmegen, the Netherlands.
Mol Imaging Biol. 2022 Feb;24(1):71-81. doi: 10.1007/s11307-021-01653-6. Epub 2021 Oct 20.
Isoflurane (ISO) is the most commonly used preclinical inhalation anesthetic. This is a problem in F MRI of fluorine contrast agents, as ISO signals cause artifacts that interfere with unambiguous image interpretation and quantification; the two most attractive properties of heteronuclear MRI. We aimed to avoid these artifacts using MRI strategies that can be applied by any pre-clinical researcher.
Three strategies to avoid ISO chemical shift displacement artifacts (CSDA) in F MRI are described and demonstrated with measurements of F-containing agents in phantoms and in vivo (n = 3 for all strategies). The success of these strategies is compared to a standard Rapid Acquisition with Relaxation Enhancement (RARE) sequence, with phantom and in vivo validation. ISO artifacts can successfully be avoided by (1) shifting them outside the region of interest using a narrow signal acquisition bandwidth, (2) suppression of ISO by planning a frequency-selective suppression pulse before signal acquisition or by (3) preventing ISO excitation with a 3D sequence with a narrow excitation bandwidth.
All three strategies result in complete ISO signal avoidance (p < 0.0001 for all methods). Using a narrow acquisition bandwidth can result in loss of signal to noise ratio and distortion of the image, and a frequency-selective suppression pulse can be incomplete when B-inhomogeneities are present. Preventing ISO excitation with a narrow excitation pulse in a 3D sequence yields the most robust results (relative SNR 151 ± 28% compared to 2D multislice methods, p = 0.006).
We optimized three easily implementable methods to avoid ISO signal artifacts and validated their performance in phantoms and in vivo. We make recommendation on the parameters that pre-clinical studies should report in their method section to make the used approach insightful.
异氟烷(ISO)是最常用于临床前吸入麻醉的药物。这在氟代造影剂的 F 磁共振成像(MRI)中是一个问题,因为 ISO 信号会产生伪影,干扰明确的图像解释和定量;这是异核 MRI 的两个最吸引人的特性。我们旨在通过任何临床前研究人员都可以应用的 MRI 策略来避免这些伪影。
描述了三种避免 F MRI 中 ISO 化学位移位移伪影(CSDA)的策略,并通过在体模和体内对含 F 试剂的测量来证明(所有策略的 n=3)。这些策略的成功与标准的快速采集弛豫增强(RARE)序列进行了比较,进行了体模和体内验证。可以通过以下三种策略成功避免 ISO 伪影:(1)使用窄信号采集带宽将其移到感兴趣区域之外;(2)在信号采集前规划频率选择抑制脉冲以抑制 ISO,或(3)使用具有窄激发带宽的 3D 序列防止 ISO 激发。
所有三种策略都导致完全避免 ISO 信号(所有方法的 p<0.0001)。使用窄采集带宽会导致信号噪声比降低和图像失真,并且当存在 B 不均匀性时,频率选择抑制脉冲可能不完全。在 3D 序列中使用窄激发脉冲防止 ISO 激发可获得最稳健的结果(与 2D 多切片方法相比,相对 SNR 为 151±28%,p=0.006)。
我们优化了三种易于实现的方法来避免 ISO 信号伪影,并在体模和体内验证了它们的性能。我们建议临床前研究在其方法部分中应报告哪些参数,以便了解所使用的方法。