Suppr超能文献

功能磁共振成像策略减少异氟醚在体内图像的伪影。

F MRI Imaging Strategies to Reduce Isoflurane Artifacts in In Vivo Images.

机构信息

Tumor Immunology Lab, Radboud Institute for Molecular Life Sciences, Radboudumc, Nijmegen, the Netherlands.

Department of Medical Imaging, Radboudumc, Nijmegen, the Netherlands.

出版信息

Mol Imaging Biol. 2022 Feb;24(1):71-81. doi: 10.1007/s11307-021-01653-6. Epub 2021 Oct 20.

Abstract

PURPOSE

Isoflurane (ISO) is the most commonly used preclinical inhalation anesthetic. This is a problem in F MRI of fluorine contrast agents, as ISO signals cause artifacts that interfere with unambiguous image interpretation and quantification; the two most attractive properties of heteronuclear MRI. We aimed to avoid these artifacts using MRI strategies that can be applied by any pre-clinical researcher.

PROCEDURES

Three strategies to avoid ISO chemical shift displacement artifacts (CSDA) in F MRI are described and demonstrated with measurements of F-containing agents in phantoms and in vivo (n = 3 for all strategies). The success of these strategies is compared to a standard Rapid Acquisition with Relaxation Enhancement (RARE) sequence, with phantom and in vivo validation. ISO artifacts can successfully be avoided by (1) shifting them outside the region of interest using a narrow signal acquisition bandwidth, (2) suppression of ISO by planning a frequency-selective suppression pulse before signal acquisition or by (3) preventing ISO excitation with a 3D sequence with a narrow excitation bandwidth.

RESULTS

All three strategies result in complete ISO signal avoidance (p < 0.0001 for all methods). Using a narrow acquisition bandwidth can result in loss of signal to noise ratio and distortion of the image, and a frequency-selective suppression pulse can be incomplete when B-inhomogeneities are present. Preventing ISO excitation with a narrow excitation pulse in a 3D sequence yields the most robust results (relative SNR 151 ± 28% compared to 2D multislice methods, p = 0.006).

CONCLUSION

We optimized three easily implementable methods to avoid ISO signal artifacts and validated their performance in phantoms and in vivo. We make recommendation on the parameters that pre-clinical studies should report in their method section to make the used approach insightful.

摘要

目的

异氟烷(ISO)是最常用于临床前吸入麻醉的药物。这在氟代造影剂的 F 磁共振成像(MRI)中是一个问题,因为 ISO 信号会产生伪影,干扰明确的图像解释和定量;这是异核 MRI 的两个最吸引人的特性。我们旨在通过任何临床前研究人员都可以应用的 MRI 策略来避免这些伪影。

过程

描述了三种避免 F MRI 中 ISO 化学位移位移伪影(CSDA)的策略,并通过在体模和体内对含 F 试剂的测量来证明(所有策略的 n=3)。这些策略的成功与标准的快速采集弛豫增强(RARE)序列进行了比较,进行了体模和体内验证。可以通过以下三种策略成功避免 ISO 伪影:(1)使用窄信号采集带宽将其移到感兴趣区域之外;(2)在信号采集前规划频率选择抑制脉冲以抑制 ISO,或(3)使用具有窄激发带宽的 3D 序列防止 ISO 激发。

结果

所有三种策略都导致完全避免 ISO 信号(所有方法的 p<0.0001)。使用窄采集带宽会导致信号噪声比降低和图像失真,并且当存在 B 不均匀性时,频率选择抑制脉冲可能不完全。在 3D 序列中使用窄激发脉冲防止 ISO 激发可获得最稳健的结果(与 2D 多切片方法相比,相对 SNR 为 151±28%,p=0.006)。

结论

我们优化了三种易于实现的方法来避免 ISO 信号伪影,并在体模和体内验证了它们的性能。我们建议临床前研究在其方法部分中应报告哪些参数,以便了解所使用的方法。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a85b/8760223/2dc292b0aca6/11307_2021_1653_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验