新生儿复苏时 NeoPuff T 型复苏器输送的潮气量和压力。

Tidal volumes and pressures delivered by the NeoPuff T-piece resuscitator during resuscitation of term newborns.

作者信息

Bjorland Peder Aleksander, Ersdal Hege Langli, Haynes Joanna, Ushakova Anastasia, Øymar Knut, Rettedal Siren Irene

机构信息

Department of Paediatrics, Stavanger University Hospital, Stavanger, Norway; Department of Clinical Science, University of Bergen, Bergen, Norway.

Department of Anaesthesia, Stavanger Univesity Hospital, Stavanger, Norway; Faculty of Health Sciences, University of Stavanger, Stavanger, Norway.

出版信息

Resuscitation. 2022 Jan;170:222-229. doi: 10.1016/j.resuscitation.2021.12.006. Epub 2021 Dec 13.

Abstract

AIM

T-piece resuscitators are commonly used for respiratory support during newborn resuscitation. This study aimed to describe delivered pressures and tidal volumes when resuscitating term newborns immediately after birth, using the NeoPuff T-piece resuscitator.

METHOD

Observational study from June 2019 through March 2021 at Stavanger University Hospital, Norway, including term newborns ventilated with a T-piece resuscitator after birth, with consent to participate. Ventilation parameters of the first 100 inflations from each newborn were recorded by respiration monitors and divided into an early (inflation 1-20) and a late (inflation 21-100) phase.

RESULTS

Of the 7730 newborns born, 232 term newborns received positive pressure ventilation. Of these, 129 newborns were included. In the early and the late phase, the median (interquartile range) peak inflating pressure was 30 (28-31) and 30 (27-31) mbar, and tidal volume was 4.5 (1.6-7.8) and 5.7 (2.2-9.8) ml/kg, respectively. Increased inflation times were associated with an increase in volume before plateauing at an inflation time of 0.41 s in the early phase and 0.50 s in the late phase. Inflation rates exceeding 32 per minute in the early phase and 41 per minute in the late phase were associated with lower tidal volumes.

CONCLUSION

There was a substantial variation in tidal volumes despite a relatively stable peak inflating pressure. Delivered tidal volumes were at the lower end of the recommended range. Our results indicate that an inflation time of approximately 0.5 s and rates around 30-40 per minute are associated with the highest delivered tidal volumes.

摘要

目的

T形管复苏器常用于新生儿复苏时的呼吸支持。本研究旨在描述使用NeoPuff T形管复苏器在足月儿出生后立即进行复苏时所输送的压力和潮气量。

方法

2019年6月至2021年3月在挪威斯塔万格大学医院进行的观察性研究,纳入出生后使用T形管复苏器进行通气且同意参与的足月儿。通过呼吸监测仪记录每个新生儿前100次充气的通气参数,并分为早期(第1 - 20次充气)和晚期(第21 - 100次充气)阶段。

结果

在出生的7730名新生儿中,232名足月儿接受了正压通气。其中,129名新生儿被纳入研究。在早期和晚期阶段,中位(四分位间距)峰值充气压力分别为30(28 - 31)mbar和30(27 - 31)mbar,潮气量分别为4.5(1.6 - 7.8)ml/kg和5.7(2.2 - 9.8)ml/kg。充气时间增加与潮气量增加相关,在早期阶段充气时间为0.41秒、晚期阶段为0.50秒时达到平台期。早期阶段充气频率超过每分钟32次、晚期阶段超过每分钟41次与较低的潮气量相关。

结论

尽管峰值充气压力相对稳定,但潮气量仍存在较大差异。所输送的潮气量处于推荐范围的下限。我们的结果表明,充气时间约为0.5秒、频率约为每分钟30 - 40次时,所输送的潮气量最高。