Suppr超能文献

数字测压法测量腰椎穿刺期间的脑脊液压力。

Digital manometry to measure cerebrospinal fluid pressure during lumbar puncture.

机构信息

Department of Radiology, 213852Duke University Medical Center, Durham, NC, USA.

出版信息

Neuroradiol J. 2022 Aug;35(4):486-489. doi: 10.1177/19714009211055190. Epub 2021 Dec 24.

Abstract

RATIONALE AND OBJECTIVES

Accurate cerebrospinal fluid (CSF) pressure measurements are critical for diagnosis and treatment of pathologic processes involving the central nervous system. Measuring opening CSF pressure using an analog device takes several minutes, which can be burdensome in a busy practice. The purpose of this study was to compare accuracy of a digital pressure measurement device with analog manometry, the reference gold standard. Secondary purpose included an assessment of possible time savings.

MATERIALS AND METHODS

This study was a retrospective, cross-sectional investigation of 71 patients who underwent image-guided lumbar puncture (LP) with opening CSF pressure measurement at a single institution from June 2019 to September 2019. Exclusion criteria were examinations without complete data for both the digital and analog measurements or without recorded needle gauge. All included LPs and CSF pressures were measured with the patient in the left lateral decubitus position, legs extended. Acquired data included (1) digital and analog CSF pressures and (2) time required to measure CSF pressure.

RESULTS

A total of 56 procedures were analyzed in 55 patients. There was no significant difference in mean CSF pressures between devices: 22.5 cm H2O digitally vs 23.1 analog ( = .7). Use of the digital manometer resulted in a time savings of 6 min (438 s analog vs 78 s digital, < .001).

CONCLUSION

Cerebrospinal fluid pressure measurements obtained with digital manometry demonstrate comparable accuracy to the reference standard of analog manometry, with an average time savings of approximately 6 min per case.

摘要

背景与目的

准确的脑脊液(CSF)压力测量对于涉及中枢神经系统的病理过程的诊断和治疗至关重要。使用模拟设备测量开放 CSF 压力需要几分钟的时间,这在繁忙的实践中可能会带来负担。本研究的目的是比较数字压力测量设备与模拟测压法的准确性,模拟测压法是参考金标准。次要目的包括评估可能的节省时间。

材料与方法

这是一项回顾性、横断面研究,纳入了 2019 年 6 月至 9 月在一家机构进行的 71 例影像学引导下腰椎穿刺(LP)伴开放 CSF 压力测量的患者。排除标准为数字和模拟测量均无完整数据或无记录的针规的检查。所有纳入的 LP 和 CSF 压力均在患者左侧卧位、腿部伸直时进行测量。获得的数据包括(1)数字和模拟 CSF 压力和(2)测量 CSF 压力所需的时间。

结果

共分析了 55 例患者的 56 个手术。两种设备的平均 CSF 压力无显著差异:数字为 22.5 cm H2O,模拟为 23.1 cm H2O(=0.7)。使用数字测压计可节省 6 分钟的时间(模拟为 438 秒,数字为 78 秒,<0.001)。

结论

数字测压计测量的 CSF 压力与模拟测压法的参考标准具有相当的准确性,平均每个病例可节省约 6 分钟的时间。

相似文献

1
Digital manometry to measure cerebrospinal fluid pressure during lumbar puncture.
Neuroradiol J. 2022 Aug;35(4):486-489. doi: 10.1177/19714009211055190. Epub 2021 Dec 24.
2
Comparison Of Digital Manometer And Water Column Manometer Pressures Measurements During Lumbar Puncture.
Med Devices (Auckl). 2019 Oct 29;12:451-458. doi: 10.2147/MDER.S225757. eCollection 2019.
7
Simple detecting of elevated ICP through liquor flow after lumbar puncture.
Am J Emerg Med. 2024 Jan;75:128-130. doi: 10.1016/j.ajem.2023.10.029. Epub 2023 Oct 30.
8
Cerebrospinal fluid outflow resistance as a diagnostic marker of spontaneous cerebrospinal fluid leakage.
J Neurosurg Spine. 2017 Aug;27(2):227-234. doi: 10.3171/2017.1.SPINE16548. Epub 2017 Jun 2.
10
Patient position during lumbar puncture has no meaningful effect on cerebrospinal fluid opening pressure in children.
J Child Neurol. 2010 May;25(5):616-9. doi: 10.1177/0883073809359198. Epub 2010 Feb 22.

本文引用的文献

1
Comparison Of Digital Manometer And Water Column Manometer Pressures Measurements During Lumbar Puncture.
Med Devices (Auckl). 2019 Oct 29;12:451-458. doi: 10.2147/MDER.S225757. eCollection 2019.
3
Acute compartment syndrome in orthopedics: causes, diagnosis, and management.
Adv Orthop. 2015;2015:543412. doi: 10.1155/2015/543412. Epub 2015 Jan 19.
4
Cerebrospinal fluid pressure in adults.
J Neuroophthalmol. 2014 Sep;34(3):278-83. doi: 10.1097/WNO.0000000000000155.
5
Diagnostic Lumbar Puncture.
Ulster Med J. 2014 May;83(2):93-102.
6
The International Classification of Headache Disorders, 3rd edition (beta version).
Cephalalgia. 2013 Jul;33(9):629-808. doi: 10.1177/0333102413485658.
7
Position-related variability of CSF opening pressure measurements.
AJNR Am J Neuroradiol. 2013 Apr;34(4):904-7. doi: 10.3174/ajnr.A3313. Epub 2012 Oct 11.
9
Cerebrospinal fluid analysis.
Am Fam Physician. 2003 Sep 15;68(6):1103-8.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验