奥地利核医学影像诊断参考水平:一项针对成年患者应用剂量水平的全国性调查。

Diagnostic Reference Levels for nuclear medicine imaging in Austria: A nationwide survey of used dose levels for adult patients.

机构信息

Austrian National Public Health Institute (Gesundheit Österreich GmbH (GÖG)), Vienna, Austria.

QIMP Team, Center for Medical Physics and Biomedical Engineering, Medical University Vienna, Austria; Austrian Society of Nuclear Medicine and Molecular Biology (OGNMB), Vienna, Austria.

出版信息

Z Med Phys. 2022 Aug;32(3):283-295. doi: 10.1016/j.zemedi.2021.11.007. Epub 2022 Jan 20.

Abstract

PURPOSE

To assess dose levels in routine nuclear medicine (NUC) procedures in Austria as a prior to a legislative update of the National Diagnostic Reference Levels (NDRL).

METHOD

As part of a nationwide survey of common NUC-examinations between June 2019 and November 2019, data sets were collected from 33 Austrian hospitals with NUC equipment. All hospitals were asked to report the NUC imaging devices in use (model, type, year of manufacture, detector material, collimators), the standard protocol parameters for selected examinations (standard activity, collimator, average acquisition time, reconstruction type, use of time-of-flight) and to report data from 10 representative examinations (e.g. injected activity, weight), incl. the most common NUC-examinations for planar imaging/SPECT and PET. Median/mean values for injected activity were calculated and compared to current Austrian and international NDRL. A Pearson correlation coefficient was computed comparing different variables.

RESULTS

In total, all 33 hospitals (100% response rate) reported data for this study for 60 SPECT devices, 21 PET/CT devices and 23 scintigraphy devices. Fixed activity values for scintigraphy/SPECT and PET were employed by about 90% and 56% of the hospitals, respectively. The most widely performed examinations for scintigraphy/SPECT are bone imaging, thyroid imaging, renal imaging (with MAG3/EC) and lung perfusion imaging (in 88% of the hospitals) and F-18 FDG-PET studies for oncology indications (in 100% of the hospitals). Significant correlations were found for patient weight and injected activity (scintigraphy/SPECT), use of iterative reconstruction and injected activity (PET) as well as size of field-of-view and injected activity (PET).

CONCLUSIONS

The reported injected activity levels were comparable to those in other countries. However, for procedures for which NDRL exist, deviations in injected activities of >20% compared to the NDRL were found. These deviations are assumed to result mainly from advances in technology but also from deviations between NDRL and prescribed activities as given in the information leaflets of the radiopharmaceuticals.

摘要

目的

在对国家诊断参考水平(NDRL)进行立法更新之前,评估奥地利常规核医学(NMU)程序中的剂量水平。

方法

作为 2019 年 6 月至 11 月期间全国性常见 NMU 检查调查的一部分,从 33 家拥有 NMU 设备的奥地利医院收集了数据集。所有医院都被要求报告正在使用的 NMU 成像设备(型号、类型、制造年份、探测器材料、准直器)、选定检查的标准协议参数(标准活动、准直器、平均采集时间、重建类型、使用飞行时间),并报告 10 项代表性检查的数据(例如,注射活动量、体重),包括平面成像/SPECT 和 PET 的最常见 NMU 检查。计算注射活动的中位数/平均值,并与当前奥地利和国际 NDRL 进行比较。计算了不同变量之间的皮尔逊相关系数。

结果

共有 33 家医院(100%回复率)报告了这项研究的 60 个 SPECT 设备、21 个 PET/CT 设备和 23 个闪烁照相设备的数据。大约 90%和 56%的医院分别使用固定的闪烁照相/SPECT 和 PET 活动值。闪烁照相/SPECT 最广泛开展的检查是骨成像、甲状腺成像、肾脏成像(使用 MAG3/EC)和肺灌注成像(在 88%的医院)以及 F-18 FDG-PET 研究用于肿瘤适应症(在 100%的医院)。发现患者体重和注射活动(闪烁照相/SPECT)、迭代重建的使用和注射活动(PET)以及视野大小和注射活动(PET)之间存在显著相关性。

结论

报告的注射活动水平与其他国家相当。然而,对于存在 NDRL 的程序,与 NDRL 相比,注射活动的偏差超过 20%。这些偏差主要归因于技术的进步,但也归因于放射性药物信息传单中规定的活动与 NDRL 之间的偏差。