GripAble Limited, London, UK.
Bioengineering Department Imperial College of Science, Technology and Medicine, London, UK.
BMC Musculoskelet Disord. 2022 Jan 24;23(1):80. doi: 10.1186/s12891-022-05026-0.
Maximum grip strength (MGS) is a reliable biomarker of overall health and physiological well-being. Therefore, an accurate and reliable measurement device is vital for ensuring the validity of the MGS assessment. This paper presents GripAble, a mobile hand grip device for the assessment of MGS. GripAble's performance was evaluated using an inter-instrument reliability test against the widely used Jamar PLUS+ dynamometer.
MGS data from sixty-three participants (N = 63, median (IQR) age = 29.0 (29.5) years, 33 M/30 F) from both hands using GripAble and Jamar PLUS+ were collected and compared. Intraclass correlation (ICC), regression, and Bland and Altman analysis were performed to evaluate the inter-instrument reliability and relationship in MGS measurements between GripAble and Jamar PLUS+ .
GripAble demonstrates good-to-excellent inter-instrument reliability to the Jamar PLUS+ with ICC = 0.906 (95% CI [0.87-0.94]). GripAble's MGS measurement is equivalent to 69% (95% CI [0.67-0.71]%) of Jamar PLUS+'s measurement. There is a proportional difference in mean MGS between the two devices, with the difference in MGS between GripAble and Jamar PLUS+ increasing with MGS.
The GripAble is a reliable tool for measuring grip strength. However, the MGS readings from GripAble and Jamar PLUS+ should not be interchanged for serial measurements of the same patient, nor be translated directly from one device to the other. A new normative MGS data using GripAble will be collected and accessed through the software for immediate comparison to age and gender-matched subpopulations.
最大握力(MGS)是整体健康和生理福祉的可靠生物标志物。因此,准确可靠的测量设备对于确保 MGS 评估的有效性至关重要。本文介绍了 GripAble,一种用于评估 MGS 的移动手部握力设备。使用与广泛使用的 Jamar PLUS+测力计的仪器间可靠性测试来评估 GripAble 的性能。
使用 GripAble 和 Jamar PLUS+从 63 名参与者(N=63,中位数(IQR)年龄=29.0(29.5)岁,33 名男性/30 名女性)的双手收集并比较了 MGS 数据。进行了组内相关系数(ICC)、回归和 Bland 和 Altman 分析,以评估 GripAble 和 Jamar PLUS+之间 MGS 测量的仪器间可靠性和关系。
GripAble 与 Jamar PLUS+具有良好到极好的仪器间可靠性,ICC=0.906(95%CI [0.87-0.94])。GripAble 的 MGS 测量值相当于 Jamar PLUS+测量值的 69%(95%CI [0.67-0.71]%)。两种设备之间的平均 MGS 存在比例差异,GripAble 和 Jamar PLUS+之间的 MGS 差异随 MGS 增加而增加。
GripAble 是一种可靠的握力测量工具。然而,GripAble 和 Jamar PLUS+的 MGS 读数不应在同一患者的连续测量中互换,也不应直接从一种设备转换到另一种设备。将使用 GripAble 收集和访问新的正常 MGS 数据,并通过软件立即与年龄和性别匹配的亚群进行比较。