J Am Podiatr Med Assoc. 2021 Dec 15;111(6). doi: 10.7547/20-241.
The literature is scanty regarding the biomechanical effects of different thread configurations on the initial stability of ankle arthrodesis. This study aims to compare the initial stability of tibiotalar fusion site in ankle arthrodesis using cannulated screws with different thread designs.
We biomechanically tested under cyclic loading the effects of different screw combinations on the initial stability of ankle arthrodesis. A total of 28 synthetic ankle models were divided into four groups: two partially threaded cancellous screws (group A), partially and fully threaded cancellous screws (group B), a partially threaded cancellous screw with a headless compression screw (group C), and a fully threaded cancellous screw and a headless compression screw (group D). Biomechanical variables including ultimate failure load, initial stiffness, ultimate stiffness, and failure angulation were analyzed.
There were no differences in any of the biomechanical variables among the four groups (P = .41 for ultimate failure load, P = .079 for initial stiffness, P = .084 for ultimate stiffness, and P = .937 for failure angulation).
Combinations of different cannulated screws showed similar results in terms of the stability and stiffness of the tibiotalar fusion site.
关于不同螺纹构型对踝关节融合初始稳定性的生物力学影响,文献报道较少。本研究旨在比较使用不同螺纹设计的空心螺钉固定踝关节融合部位的初始稳定性。
我们通过循环加载对不同螺钉组合对踝关节融合初始稳定性的影响进行生物力学测试。共 28 个合成踝关节模型分为四组:两组部分螺纹松质骨螺钉(组 A)、部分和全螺纹松质骨螺钉(组 B)、带无头加压螺钉的部分螺纹松质骨螺钉(组 C)和全螺纹松质骨螺钉和无头加压螺钉(组 D)。分析生物力学变量,包括最终失效载荷、初始刚度、最终刚度和失效角度。
四组间任何生物力学变量均无差异(最终失效载荷 P =.41,初始刚度 P =.079,最终刚度 P =.084,失效角度 P =.937)。
不同空心螺钉的组合在距骨 - 胫骨融合部位的稳定性和刚度方面具有相似的结果。