Radovanovic Mirjana, Jones Graham, Day Richard O, Galettis Peter, Norris Ross L G
St Vincent's Clinical School, University of New South Wales, Darlinghurst, NSW, Australia.
SydPath, St Vincent's Hospital Sydney, Darlinghurst, NSW, Australia.
J Mass Spectrom Adv Clin Lab. 2022 Apr 26;24:57-64. doi: 10.1016/j.jmsacl.2022.04.002. eCollection 2022 Apr.
Utilising stable isotope labelled internal standards (SIL-IS) in quantitative LC-MS/MS drug analysis is the most widely used approach to normalise for variability during sample quantification processes. However, compounds containing atoms such as Sulphur, Chlorine or Bromine, could potentially cause cross-signal contribution to the SIL-IS from the naturally occurring isotopes, resulting in non-linear calibration curves. A simple, novel method of mitigating the effect is presented here. It entails monitoring of a less abundant SIL-IS isotope, as the precursor ion, of a mass that has no/minimal isotopic contribution from the analyte isotopes.
Experiments were conducted on two LC-MS/MS analysers: Waters Xevo TQ-S and Shimadzu 8050. Flucloxacillin (FLX) was used as an example. Two transitions were selected for FLX (/ 454 → 160 → 295) and one for each of the SIL-IS isotopes (/ 458 → 160 for the isotope 457 g/mol and / 460 → 160 for the isotope 459 g/mol). Assay biases were assessed at three SIL-IS concentrations: 0.7, 7 and 14 mg/L for each isotope.
When using the SIL-IS isotope / 458 → 160 at a concentration of 0.7 mg/L, biases were up to 36.9 % on both instruments. Increasing the SIL-IS concentration to 14 mg/L, reduced the bias to 5.8 %. Using the less abundant isotope, / 460 → 160, resulted in biases of 13.9 % at an SIL-IS concentration of 0.7 mg/L.
Applying this method will mitigate cross-signal contribution from the analyte isotopes to the corresponding SIL-IS, minimise the use of SIL-IS, and, thereby, reduce overall cost.
在定量液相色谱-串联质谱(LC-MS/MS)药物分析中,使用稳定同位素标记内标(SIL-IS)是样品定量过程中用于标准化变异性的最广泛使用的方法。然而,含有硫、氯或溴等原子的化合物可能会导致天然存在的同位素对SIL-IS产生交叉信号贡献,从而导致校准曲线非线性。本文提出了一种简单、新颖的减轻这种影响的方法。该方法需要监测一种丰度较低的SIL-IS同位素,作为前体离子,其质量没有/最小程度受到分析物同位素的同位素贡献。
在两台LC-MS/MS分析仪上进行实验:沃特世Xevo TQ-S和岛津8050。以氟氯西林(FLX)为例。为FLX选择了两个跃迁(/ 454 → 160 → 295),为每个SIL-IS同位素选择了一个跃迁(同位素457 g/mol的/ 458 → 160和同位素459 g/mol的/ 460 → 160)。在三种SIL-IS浓度下评估测定偏差:每种同位素分别为0.7、7和14 mg/L。
当使用浓度为0.7 mg/L的SIL-IS同位素/ 458 → 160时,两台仪器上的偏差高达36.9%。将SIL-IS浓度提高到14 mg/L,偏差降低到5.8%。使用丰度较低的同位素/ 460 → 160,在SIL-IS浓度为0.7 mg/L时偏差为13.9%。
应用该方法将减轻分析物同位素对相应SIL-IS的交叉信号贡献,最大限度地减少SIL-IS的使用,从而降低总体成本。