超声多普勒设备声学输出的一项调查。

A survey of the acoustic output of ultrasonic Doppler equipment.

作者信息

Duck F A, Starritt H C, Anderson S P

出版信息

Clin Phys Physiol Meas. 1987 Feb;8(1):39-49. doi: 10.1088/0143-0815/8/1/002.

Abstract

Measurements of the acoustic output generated by a variety of clinical ultrasonic Doppler instruments have been carried out. The instruments surveyed include continuous wave Doppler units for cardiovascular investigations, fetal monitors, stand-alone pulsed Doppler equipment and 'duplex' scanners working in Doppler mode. Acoustic measurements have been made using a calibrated PVDF membrane hydrophone, and a milliwatt radiation force balance. Almost all the pulsed Doppler and duplex systems investigated could generate spatial-peak, temporal-average intensities in water which exceeded 100 mW cm-2, with a maximum of 825 mW cm-2 measured. These intensities were also reached by some continuous-wave Doppler systems, particularly those incorporated into duplex scanning systems. Fetal monitoring equipment was found to operate typically at lower intensities. Doppler units have been found to show a very wide variation in pulse length, repetition frequency and spatial peak pressure, and a wider range of pulse average intensities than previously reported. Units were found to vary considerably in their control of total acoustic power whilst operating parameters such as gate width and range were altered.

摘要

已对多种临床超声多普勒仪器产生的声输出进行了测量。所调查的仪器包括用于心血管检查的连续波多普勒装置、胎儿监护仪、独立的脉冲多普勒设备以及工作在多普勒模式下的“双功”扫描仪。已使用校准的聚偏二氟乙烯(PVDF)膜水听器和毫瓦辐射力天平进行了声学测量。几乎所有被研究的脉冲多普勒和双功系统在水中产生的空间峰值、时间平均强度都超过了100 mW/cm²,测量到的最大值为825 mW/cm²。一些连续波多普勒系统,特别是那些集成到双功扫描系统中的系统,也达到了这些强度。发现胎儿监护设备通常在较低强度下运行。已发现多普勒装置在脉冲长度、重复频率和空间峰值压力方面有非常大的变化,并且脉冲平均强度范围比以前报道的更广。发现各装置在改变诸如门宽和量程等操作参数时,对总声功率的控制有很大差异。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验