在两种模拟临床条件下Dentaport ZX、Rootor和E-Pex Pro电子根尖定位仪准确性的比较:一项研究。

Comparison of Accuracy of Dentaport ZX, Rootor and E-Pex Pro Electronic Apex Locators in Two Simulated Clinical Conditions: An Study.

作者信息

Mohan Malavika, Verma Meenakshi R, Jain Ashish K, Rao Rahul D, Yadav Priyanka, Agrawal Sonal

机构信息

Department of Conservative Dentistry and Endodontics, Bharati Vidyapeeth Dental College and Hospital, Navi Mumbai, Maharashtra, India.

出版信息

J Conserv Dent. 2022 Jan-Feb;25(1):58-62. doi: 10.4103/jcd.jcd_557_21. Epub 2022 May 2.

Abstract

BACKGROUND

The aim of this study was to determine the accuracy in measuring the working length (WL) using Dentaport ZX, Rootor, and a newly introduced budget friendly electronic apex locator (EAL), E-Pex Pro in two commonly simulated clinical conditions (in the presence of irrigant and blood).

MATERIALS AND METHODOLOGY

Eighty-eight single-rooted premolars were randomly assigned to two groups according to simulated clinical conditions: Group 1: Presence of irrigant (5% NaOCl) and Group 2: Presence of blood. WL was determined with all three EAL and was then compared with actual length (AL) of the tooth, which was measured using Vision Inspection System. The difference between the length measured by EAL and AL was tabulated and statistically analyzed using one-way analysis of variance (ANOVA) with Dunnett's test. All testing was done using two-sided tests at alpha 0.05 (95% confidence level). Thus, the criteria for rejecting the null hypothesis were "P < 0.05."

RESULTS

Measurement using Dentaport ZX, Rootor, and E-Pex Pro had an accuracy of 99.79%, 99.69%, and 99.64%, respectively, in Group 1 and 99.95%, 99.7%, and 99.74%, respectively, in Group 2. ANOVA revealed that the mean error value is least for Dentaport ZX followed by Rootor and then E-Pex pro EAL.

CONCLUSION

Dentaport ZX gave better results both in the presence of NaOCl and blood followed by Rootor and E-Pex Pro EALs.

摘要

背景

本研究的目的是确定在两种常见的模拟临床条件下(存在冲洗液和血液),使用登士柏ZX、Rootor以及一种新推出的经济实惠的电子根尖定位仪(EAL)E-Pex Pro测量工作长度(WL)的准确性。

材料与方法

根据模拟临床条件,将88颗单根前磨牙随机分为两组:第1组:存在冲洗液(5%次氯酸钠);第2组:存在血液。使用所有三种EAL确定WL,然后将其与使用视觉检查系统测量的牙齿实际长度(AL)进行比较。将EAL测量的长度与AL之间的差异制成表格,并使用单因素方差分析(ANOVA)和Dunnett检验进行统计分析。所有测试均使用双侧检验,显著性水平为0.05(95%置信水平)。因此,拒绝原假设的标准是“P < 0.05”。

结果

在第1组中,使用登士柏ZX、Rootor和E-Pex Pro测量的准确率分别为99.79%、99.69%和99.64%;在第2组中,分别为99.95%、99.7%和99.74%。方差分析显示,登士柏ZX的平均误差值最小,其次是Rootor,然后是E-Pex pro EAL。

结论

在存在次氯酸钠和血液的情况下,登士柏ZX的结果优于Rootor和E-Pex Pro EAL。